MAINTENANCE FUNCTIONS Work Volume Measurement Using Flow Master
8. Press the ENT key.
9. Press the START key. This will begin by testing a non-leaking item. (Volume testing start)
10. It then begins testing a leaking item. (Volume testing start)
11. It will indicate the volume following the test.
12. Press the START key. (This will record the internal volume value)
13. Press the STOP key twice. (This will return to the initial screen)
OK
*******
ml/m
Exhaust
Test results are not relevant.
OK
*******
ml/m
Exhaust
Test results are not relevant.
Work Volume 7.093E1 ml
Startkey
→
Data update
Stopkey
→
Cancel
Maint
1/2P
*
Volume Measurement
I/Ocheck
SelfCheck
StabilityChck
SoftVersion
Manual
+000
kPa
Ready GroupNo 00
When the volume measurement
result is not stable even after
repeated measurement, perform
the following procedure:
(1)Extend the exhaust time.
(2)Exchange the flow master with
a large leak amount.