BASIC OPERATIONS SETTING PROCEDURES
■
Ordinary Testing Procedure
1. Press the START key
.
(Testing starts)
(If using a Remote, input the START signal from the REMOTE connector.)
2. The test result will be output when the testing is concluded.
(
*
1) For operations 1~10, once the settings are set, there is no need to repeat that process for each test.
When setting values are changed, the first test will always be a mandatory fitting testing.
(*
2
)
If you choose fitting set 0, 1 or 2, and then alter a setting item which
influences the measurement conditions, the first measurement (The 1
st
time
measurement) will be a mandatory fitting measurement. The setting items
which influence measurement conditions are: Charge 1, Charge 2, Balance 1, Balance 2,
Detect, F Detect, Lead P., Test P., and Fitting Set.
OK
1.000
ml/m
Exhaust
GroupNo 00
・
Is the AIR SUPPLY(Test Pressure) being supplied?
・
Is the A.P SUP (Pilot Valve Operation Pressure) being supplied?
・
Is the work bowl valve opened?
・
Is the work connected to the work port?
・
Are each of the setting values correct?
Preparations
Test Value
Judgment
Result
Manual
-
0000Pa
-
0
+
Detect
GroupNo 00
Содержание FL-601 Series
Страница 10: ...9 OUTLINE ...
Страница 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Страница 18: ...17 PRINCIPLE ...
Страница 24: ...23 POWER ON PROCEDURE ...
Страница 50: ...49 BASIC OPERATION ...
Страница 59: ...58 TESTING OPERATION ...
Страница 68: ...67 MAINTNANCE FUNCTIONS ...
Страница 92: ...91 OPTIONAL EQUIPMENT ...
Страница 108: ...111 SUPPLEMENT ...
Страница 112: ...111 SUPPLEMENT ...
Страница 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Страница 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...