POWER ON PROCEDURE INSTALLATION INSTRUCTIONS
■
INSTALLATION INSTRUCTIONS
Air leak testers detect leaks by detecting precise variations in pressure (air condition.)
Therefore, it requires handling instructions that are different from standard precision equipment.
Tester must be in a stable position.
・
Be sure to install the tester on a level surface.
Vibrations or earthquakes may cause the tester to fall and cause injury to
operators.
WARNING
Do not place anything on top of the tester.
・
Do not place anything on top of the tester.
Falling objects may result in injury.
Do not install the tester in the following locations:
・
High moisture or dusty areas.
・
Areas in which oil and water can cause contamination.
・
Areas directly exposed to sunlight.
・
Outdoors
This may result in fire or electric shock.
・
Areas with obstructions or places in which the tester ’s fanning
system may be blocked.
The tester may overheat and may cause a fire.
CAUTION
Moving the Tester:
・
Be sure to shut off the power, unplug any cords, shut off the
pneumatic source and remove any piping.
・
Be sure to carry the tester in a stable manner.
(Do not carry the tester by the manifold handle.)
This may cause damage to the tester or to its casing and may
result in injury.
○
OPERATION ENVIRONMENT
・
For best results and accuracy, it is recommended that the operation temperature
and environmental humidity conditions are 23
±
5
℃
, 55
±
10%RH
・
If there is an abrupt temperature change, noise will be added to the test data,
producing unreliable results.
・
Transferring the tester rapidly from high to low temperature locations and/or
humidity locations may cause condensation. When such cases cannot be
avoided, leave the tester in the environment for at least an hour in order for it
to adjust to the surrounding temperature.
FL-601
・
Do not install the tester near windows or other locations
affected by direct sunlight.
・
Ambient Temperature: 0~40
℃
・
Ambient Humidity: 45~80% RH with no condensation.
・
Install the tester where it will not receive any influences
from temperature or humidity changes, such as at the
center of the facility.
FL-601
・
Do not install the tester where it will receive direct
influence from wind or blown air.
・
Install the tester in a location where air from air
conditioning units, etc will not affect the tester’s
performance.
Provide a cover for the tester and/or other fixtures if
necessary.
WARNI
NG
Содержание FL-601 Series
Страница 10: ...9 OUTLINE ...
Страница 17: ...16 OUTLINE ADDITIONAL EQUIPMENT MEMO ...
Страница 18: ...17 PRINCIPLE ...
Страница 24: ...23 POWER ON PROCEDURE ...
Страница 50: ...49 BASIC OPERATION ...
Страница 59: ...58 TESTING OPERATION ...
Страница 68: ...67 MAINTNANCE FUNCTIONS ...
Страница 92: ...91 OPTIONAL EQUIPMENT ...
Страница 108: ...111 SUPPLEMENT ...
Страница 112: ...111 SUPPLEMENT ...
Страница 116: ...115 SUPPLEMENT Pneumatic Circuit Diagram FL 601UL L M H1Pneumatic Circuit Diagram ...
Страница 119: ...118 SUPPLEMENT Pneumatic Circuit Diagram ChargeBypassUnit CBU 600C Exhaust Valve ...