![Fukuda FL-286 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/fukuda/fl-286/fl-286_installation-instructions-manual_2344841028.webp)
60
RS-232C CONNECTION
☆
This is an optional function.
z
The test data can be sent to an external computer and other devices using this RS-232C
output. The transmission data is as follows.
•
Transmit measurement data
•
Transmit measurement condition
•
Receive measurement condition
z
RS-232C Output Specifications.
•
Transmission speed
9600 bps
•
Start Bit
1 bit
•
Stop Bit
1 bit
•
Data Length 8 bit
•
Parity None
•
Text Encoding ASCII
•
Hardware Composition DTE end (see connector table)
•
Transmission Method
Asynchronous, half-duplex, RTS/CTS handshake
•
Connector
25 pin D-SUB
•
Transmit measurement condition
•
Receive measurement condition
z
Connector Table and Connection Example
Connector 25pin
-
PC 9 pin
Number Signal
name
Air Leak
Tester
Connection
PC 9pin
Signal
name
Number
1
FG
DCD
1
2
TxD
RxD
2
3
RxD
TxD
3
4
RTS
DTR
4
5
CTS
SG
5
6
DSR
DSR
6
7
SG
RTS
7
8
DCD
CTS
8
9~19
NO
CONNECTION
RI
9
20
DTR
21~25
NO
CONNECTION
Connector 25 pin
-
PC 25 pin
Number Signal
name Air Leak
Tester
Connection
PC 25pin
Signal
name
Number
1 FG
FG
1
2 TxD
TxD
2
3 RxD
RxD
3
4 RTS
RTS
4
5 CTS
CTS
5
6 DSR
DSR
6
7 GND
SG
7
8
~
19
NO
CONNECTION
DCD
8
20 DTR
DTR 20
21
~
25
NO
CONNECTION
RI 22