10. Troubleshooting
39
ECG
Phenomenon Cause
Solution
Check the connection
between the lead cable and
the LX-7120.
Lead cable is off.
Check the connection
between the lead cable and
the electrode.
Faulty Lead cable. Replace the ECG cable with
a new one.
Electrode is peeling
off.
Replace the electrode with a
new one.
“ELECTRODE?”
message is displayed.
Polarization
potential of the
electrode is too
high.
Replace the electrode with a
new one.
Electrode gel is dry.
Electrode is peeling
off.
Replace the electrode with a
new one.
Electric blanket is
used.
Cover the electric blanket
with a shield cover.
ECG waveform
contains noise
AC filter setting of
the receiving
monitor is OFF.
Set the AC filter up as ON.
Electrode gel is dry.
Electrode is peeling
off.
Replace the electrode with a
new one.
Respiration waveform
cannot be measured.
The positions of the
electrodes are
improper.
Attach the electrodes where
the respiration waveform can
be measured appropriately.
Содержание LX-7120
Страница 18: ...xvi Blank Page ...
Страница 22: ...1 General Description 2 Blank Page ...
Страница 24: ...2 Names of Parts and Their Functions 4 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 44: ...6 Measurement 24 Blank Page ...
Страница 50: ...7 Operation 30 Blank Page ...
Страница 54: ...8 Other Setting Items 34 Blank Page ...
Страница 56: ...9 Changing the Transmitter Channel and Group ID 36 Blank Page ...
Страница 62: ...11 Cleaning and Disinfection 42 Blank Page ...
Страница 73: ......