6. Measurement
21
Displays the detection marker when the
pacemaker pulse is detected.
Indicates the measuring lead.
Indicates the ECG waveform size
displayed on the LCD.
CAUTION
The ECG waveform size setting displayed on the LCD of the LX-7120
does not interact with the one displayed on the screen of the receiving
monitor, because the LX-7120 cannot transmit the setting information of
the waveform size to the receiving monitor. If the ECG waveform size
displayed on the screen of the receiving monitor is changed, follow the
instruction in the operation manual of the receiving monitor.
In case of the outside of the heart rate range (12 to 300bpm), 0bpm will
be displayed if 11bpm and below is measured and 300bpm will be
displayed If 300bpm and above is measured.
Indicates the remaining battery level.
For details of the battery level, refer to
page 17.
Displays the electrode check
“ELECTRODE?” message appears
when the ECG electrode is detached.