5. Respiration Monitoring
Follow the preparation of “4.ECG Monitoring” to allow the respiration
monitoring.
This respiration monitoring is performed with impedance method.
The ECG electrodes are also used for detecting the respiration. Each lead
cable specifies the electrodes to detect the respiration. For 3-electrode and 5-
electrode (chest) lead cable, the electrodes to detect the respiration are fixed
as follows. Even if lead method is switched, they are no changes.
Lead Cable
Color of Electrode
3-electrode
White (RA) and Red (LL)
4-electrode
White (RA) and Red (LL)
5-electrode (Chest)
White (RA) and Red (LL)
WARNING
Minute ventilation rate-adaptive implantable pacemakers can occasionally
interact with certain cardiac monitoring and diagnostic equipment, causing
the pacemakers to pace at their maximum programmed rate. The cardiac
monitoring and diagnostic equipment may possibly send wrong
information.
If such event occurs, please disconnect the cardiac monitoring and
diagnostic equipment, or follow the procedures described in the operation
manual of the pacemaker.
(For more details, contact FUKUDA DENSHI personnel, your institution’s
professionals, or your pacemaker distributors.)
Reference
“Minute Ventilation Rate-Adaptive Pacemakers”
FDA alerts health professionals that minute ventilation rate-adaptive
implantable pacemakers can occasionally interact with certain cardiac
monitoring and diagnostic equipment, causing pacemakers to pace at their
maximum programmed rate.
[October 14, 1998 (Letter: www .fda.gov/cdrh/safety.html) - FDA]
CAUTION
Even if the electrodes are attached on the proper positions for ECG
monitoring, it may not be always the proper ones for respiration
monitoring as well.
When a defibrillator is used during respiration monitoring, a large offset
voltage will be placed on the ECG electrodes, which may cause
interruption of monitoring for a few seconds.
Содержание LX-7120
Страница 18: ...xvi Blank Page ...
Страница 22: ...1 General Description 2 Blank Page ...
Страница 24: ...2 Names of Parts and Their Functions 4 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 44: ...6 Measurement 24 Blank Page ...
Страница 50: ...7 Operation 30 Blank Page ...
Страница 54: ...8 Other Setting Items 34 Blank Page ...
Страница 56: ...9 Changing the Transmitter Channel and Group ID 36 Blank Page ...
Страница 62: ...11 Cleaning and Disinfection 42 Blank Page ...
Страница 73: ......