4.2.1 Preparing the Mouthpiece
(1) Have the patient hold the mouthpiece in his/her mouth.
[Note]
If you choose to have the patient hold the mouthpiece after
insertion, attach the mouthpiece to the insertion portion in
advance.
Have the patient hold it promptly after insertion.
4.2.2 Insertion
(1) Give instructions a patient to lie on examining table in a
proper position according to endoscopy procedures.
(2) Unlock the bending section by turning the up-down and
left-right angulation locks in the direction of F until
they stop.
[Note]
We recommend the procedure above. However, another
procedure is also available: you can insert ultrasonic
endoscope by locking the bending section only in the
left-right direction and unlocking it in the up-down
direction.