3.5 Connecting Ultrasonic Endoscope (Attachment)
CAUTION
Set the suction pressure between 40 and 53 kPa. If the suction
pressure is too high, the ultrasonic endoscope may adhere to mucous
membrane, resulting in damage to the mucous membrane.
3.5.1 Connecting the Ultrasonic Endoscope (Processor, Light Source, Water Tank and Suction Unit)
(1) Before attaching the LG connector to the light source,
visually check that there is no dirt on the light incident
distal face and no cracks on the glass of the light incident
distal face.
(2) Insert the LG connector of the ultrasonic endoscope into
the ultrasonic endoscope socket on the light source until it
stops.
(3) Insert the video connector of the ultrasonic endoscope into
the video connector socket on the processor.
Align the mark of the connector with the one of the
connector socket, and attach the connector firmly by
turning it clockwise while slightly pushing it.
Light incident distal face
Chapter 3 Preparation and Inspection Before Use
54