1.3.2 Preventing Electrical Shock
WARNING
Insert the AC plug into a hospital grade receptacle. Not doing so may
cause an electric shock accident.
Use an electrosurgical instrument conforming to EN 60601-2-2. Refer
to the manual of the electrosurgical instrument for how to operate the
electrosurgical instrument.
Connect the electrosurgical instruments and electrosurgical generators
in accordance with each operation manual. Incorrect connection may
cause electric shock and/or burns.
CAUTION
The ultrasonic endoscope may be used in combination with
peripherals. To avoid an electric shock accident, do not use any
peripherals than the ones specified in this manual.