![Fuji Electric ZSS-7 Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/fuji-electric/zss-7/zss-7_instruction-manual_2341920099.webp)
- 92 -
INZ-TN7ZSS-E
6.11.3 Operation method
(1)
Remove the stopper at gas inlet and outlet of the check cell, and then connect with N
2
gas cylinder.
(2)
Feed N
2
gas to the check cell to obtain the measured gas concentration inside the stack.
(3)
Feed the gas, which was obtained by “6.11.2 Concentration of the gas which is fed to the check cell and
gas concentration fluctuation”, to the check cell.
(4)
After the indication value is stabled, obtain the measured value to check the discrepancy of the span
point.
(5)
Feed N
2
gas to the check cell so that the environment where N
2
gas exists inside the check cell is created.
(6)
Attach the stopper so that the measured gas component does not come into the check cell.
Содержание ZSS-7
Страница 1: ......
Страница 109: ...102 INZ TN7ZSS E APPENDIX 2 CODE SYMBOLS...
Страница 114: ...INZ TN7ZSS E 107 13 BNC cable for optical axis adjustment...
Страница 116: ......