![Fuji Electric ZSS-7 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/fuji-electric/zss-7/zss-7_instruction-manual_2341920014.webp)
INZ-TN7ZSS-E
- 7 -
AI terminal
7 8 9 10 11 12
1 2 3 4 5 6
Cross sectional area of rated cable:
AWG26-16
1 AI1+
Analog input 1 (AI1)
2 AI1–
3 AI2+
Analog input 2 (AI2)
4 AI2–
5 AI3+
Analog input 3 (AI3)
AI expansion board
required
6 AI3–
7 AI4+
Analog input 4 (AI4)
8
AI4–
9
AI5+
Analog input 5 (AI5)
10
AI5–
11
AI6+
Analog input 6 (AI6)
12
AI6–
DI/DO terminal
25 26 27 28 29 30 31 32
33
34
35
36
13 14 15 16 17 18 19 20
21
22
23
24
1 2 3 4 5
6 7 8
9
10
11
12
1 DI1
Remote average value reset (option)
2 DI1
3 DI2
Switching input for remote instantaneous value/average value (option)
4 DI2
5 DI3
Remote hold (option)
6 DI3
7
DI4
8
DI4
9 DI5
Spare
10 DI5
11 Unassigned
12 Unassigned
13 DO1
Low Light Transmission
25 DO7
14 DO1
26 DO7
15 DO2
Analyzer faulty
27 DO8
Power OFF
16 DO2
28 DO8
17 DO3
On Hold / Under Calibration
29 DO9
18 DO3
30 DO9
19 DO4
Beyond the upper/lower limits
(Ch.1)
31 DO10
20 DO4
32 DO10
21 DO5
33 DO11
Beyond the upper/lower limits
22 DO5
34 DO11
(Ch.2)
23 DO6
35 DO12
24 DO6
36 DO12
Note 1) Unassigned terminals may be connected to the internal circuit.
Do not use them as repeating terminals.
Note 2) Provide relay output of LD failure, LD temperature error, or connection error as analyzer faulty.
Note 3) Do not provide relay output of high gas temperature, air purge (low pressure), box temperature
warning, PD over range or AI under.
Note 4) Consult our sales representative to collectively provide relay output of alarm depending on the in-
stallation environment.
Note 5) All the alarms except "Connection Error" which indicates communication failure between the re-
ceiver unit and the control unit will be activated after at least five minutes from the power off.
The alarm of "Low Light Transmission" is activated after 1 minute continuation. Though it takes
6 minutes from just after the power on.
Cross sectional area of rated cable:
AWG26-16