INZ-TN7ZSS-E
- 51 -
6.2.4 Measurement of average dust concentration of the same
time as the sampling for manual dust analysis conducted
Matching the value obtained by manual analysis is required for measument of dust.
Use “Average Dust Value” screen to measure the dust concentration of the same time as sampling for manual
analysis was conducted. When the result of manual analysis is obtained at a later date, perform matching by
using the “Span Calibration” screen.
Perform matching to the value obtained by manual analysis in a state that the receiver unit and the transmit-
ter unit are attached to the stack. Calibration cell is not required.
(1) If the power is off, turn on the power.
(2) Wait more than an hour after the power is turned on to perform calibration, because it takes an hour for
warm up.
(3) Display the “Span Calibration” screen.
(4) Point the to “Component”, and press the
key.
(5) Press the
key or the
key to select Dust, and press the
key.
(6) [Ave. Dust Value] appears next to [Span Cal. Start].
(7) Press the
key or the
key to move the cursor in [Ave. Dust Value], and press the
key.
The screen then transits to “Average. Dust Value” screen.
(8) There are four methods to measure the average dust concentration.
1) Measuring the average dust concentration during the period from Start Time (pre-set value) to when
the Sampling Time (pre-set value) elapses. (After the input value is set, make sure to press Start key.)
2) Measuring the average dust concentration during the period from when [Start key] is input (arbi-
trary-set time) to when the Sampling Time (pre-set value) elapses. (Value earlier than the time dis-
played in the up right corner of the screen should be set)
3) Measuring the average dust concentration during the period from Start Time (pre-set value) to when
[Finish key] (arbitrary-set time) is input. (After the Start Time is entered, make sure to press the Start
key)
4) Measuring the average dust concentration during the period from when [Start key] (arbitrary-set time)
is input to when the [Finish key] (arbitrary-set time) is input. (Value earlier than the time displayed in
the up right corner of the screen should be set)
(9) Setting change of [Start Time] or [Sampling Time] during measurement of average dust concentration
is not reflected on the display. Press the [Finish key] first, and then press the [Start key] to reflect the
change on the display.
ENT
ENT
2006-04-01
12:10:00
Start
Finish
100.0
mg/m
3
0.000
mg/m
3
06/04/01 12:10
Min
30
06/04/01 00:00:00
2006-04-01
12:10:00
Span Cal.Start
Cal.Finish
500
mm
1000.00
mg/m
3
0025
°C
kPa
Dust
00.00
100.00
mg/m
3
Range 1
2006/09/01
Ave. Dust Value
Start Time
Sampling Time
Ave
Meas.
Last time sampling day
Average Dust Value
Component
Cal. Range
Cell Length
Cal.Gas Value
Meas.
Gas Temp.
Gas Press.
LastCal.Day
Span Calibration