23
INZ-TN5A0506-E
Relative
adderess
Registrarion
number
Data
type
Memory contents
Read out /weite in data
Default
value
45 40046
O
2
concentration alarm lower2
limit value [vol%] {Upper}
(Range 2)
46 40047
DWORD
O
2
concentration alarm lower
limit alarm value [vol%]
{Lower} (Range 2)
1 to 550000
(1digit = 0.0001 vol%)
100
47 40048
WORD
Hysteresis width of O
2
concentration alarm condition
[%] (Range 2)
0 to 20
10
AO Hold setting type [Upper]
1: Valid, 0: Invalid
0
48 40049
WORD
AO Hold selection value
{Lower}
0: 0%
1: 100%
2: last time calculated value
3: User specified
0
49
40050
WORD AO Hold setting value
0 to 100
0
50 40051
WORD
Measurement return time
[Second]
0 to 300
10
51
40052
WORD Factory default setting flag
8 bits: Key lock (0:Invalid,
1:Valid)
0
52 40053
WORD
LCD Automatic OFF time of the
backlight [Minute]
0 to 99
10
53 40054
WORD
Warm-up operation monitoring
time [Minute]
0 to 60
45
54 40055
WORD
Average time of movement
[Seconds]
0 to 60
2
55 40056
WORD
Heater control system setting
(Controlled temperature)
700 to 900
800
56 40057
WORD
Heater control system setting
(cryogenic thermometry)
0 to 300
200
57 40058
WORD
Heater control system setting
(cryogenic temperature)
0 to 60
0
58 40059
WORD
Unused
-
59 40060
WORD
4mA (0V) Adjustment value
(Analog output)
0 to 65535
5250
60 40061
WORD
20mA(1V) Adjustment value
(Analog output)
0 to 65535
27000
61 40062
Zero coefficient {Upper} (Range
1)
62 40063
DWORD
Zero coefficient {Lower} (Range
1)
0 to 4294967295
(1digit = 0.001)
1000
63 40064
Span coefficient {Upper} (Range
1)
64 40065
DWORD
Span coefficient {Lower} (Range
1)
0 to 4294967295
0
65 40066
Zero coefficient {Upper} (Range
2)
1000
66 40067
DWORD
Zero coefficient {Lower} (Range
2)
0 to 4294967295
(1digit = 0.001)