20
INZ-TN5A0506-E
7.2 Address map
Word data [read-out/write-in]: Function code [03
H
, 06
H
, 10
H
]
Relative
adderess
Registrarion
number
Data
type
Memory contents
Read out /weite in data
Default
value
Unit {Upper}
0: vol%
0
0 40001
WORD
Decimal point position of range
{Lower}
0: <0.000>
1: <00.00>
1
1 40002
WORD
Range
0 to 9999
(When the decimal point
position “00.00” is 0200 to
5000, when “0.000” is 2000
to 9999)
2500
Unit {Upper}
0: vol%
0
2 40003
WORD
Decimal position of range
{Lower}
0: <0.000>
1: <00.00>
1
3 40004
WORD
Range
0 to 9999
(When the decimal point
position “00.00” is 0200 to
5000, when “0.000” is 2000
to 9999)
2500
4 40005
WORD
Max.value/Mini.value calculation
time [time]
0 to 240
24
5 40006
WORD
Setting
flag
2 bits: Inhabitation of all
calibration (0: enable, 1:
unenable)
0
Year of automatic calibration star
{Upper}
00 to 99
99
6 40007
WORD
Month of automatic calibration
start {Lower}
1 to 12
1
Date of automatic calibration
start {Upper}
1 to 31
1
7 40008
WORD
Time of automatic calibration
start{Lower}
0 to 23
0
Minute of automatic calibration
start {Upper}
0 to 59
0
8 40009
WORD
Unused 0
0
Auto calibration cycle (Date)
{Upper}
0 to 99
7
9 40010
WORD
Auto calibration cycle (Time)
{Lower}
0 to 99
0
10 40011
Calibration span gas
concentration {Upper} (Range 1)
11 40012
DWORD
Calibration span gas
concentration {Lower} (Range 1)
00010 to 50000
(1 digit = 0.001 vol%)
206000
12 40013
Calibration span gas
concentration {Upper} (Range 1)
13 40014
DWORD
Calibration zero gas
concentration {Lower} (Range 1)
00010 to 25000
(1 digit = 0.001 vol%)
20000
Содержание ZKM