5.2 Moving to the Parameter Setup Mode
5.3 Data Lists and Default Settings
5.4.5 Method of Setting Linearization Data
The linearization function means a function that changes the slope of straight
lines in the relationship between the input and indication by correcting the rela-
tions at arbitrary points.
Linearization data are set using the input value (indicated
value before correction) and the output value (indicated value after correction) at
each arbitrary point.
Note: The setup conditions are N-1 < N-2 … N-15 < N-16.
6. Control Functions
6.1 Hold Function
The Hold function temporarily retains the indication. The hold function is enabled
by shortcircuiting the HOLD and COM terminals or setting both terminals to the same
voltage level. As a result the display unit retains the indication given at that moment.
6.2 Digital Zero Function
The Digital Zero function zeros the indication given at an arbitrary timing.
Thereafter, the function shows the amount of change from the point of zero-
ing. However, this function serves as an indication resetting function for a
frequency measuring unit. Thus, the Digital Zero function can be used to reset
the indication when there is no input signal at all.
Note that, the on/off control of the Digital Zero function can be achieved by
means of terminal control or front panel keys. In the case of terminal control,
the Digital Zero function is turned on by shortcircuiting the DZ and COM termi-
nals or setting both terminals to the same voltage level. The indication at that
moment is zeroed. In the case of control with the front panel keys, hold down
the Mode key and press the Increment key for about 1 second to zero the indi-
cation at that moment.
Note: Operation with the control terminals takes priority over operation
with the front panel keys. The Digital Zero function is disabled if
the control terminals are made to go through the off-on-off se-
quence with the function enabled by means of the front panel keys.
6.3 Peak Hold Function
The Peak Hold function retains one of the maximum (peak hold)/minimum
(valley hold)/maximum - minimum (peak-valley hold) values and provides
output for that value. Selection from these values is made using the condition
data. The peak hold function is enabled by shortcircuiting the PH and COM
terminals or setting both terminals to the same voltage level.
7.
Output Function
7.1 Comparison Output Function
The FD5000 series of unit meters is designed so that the two judgment values
HI and LO can be set for the measured (indicated) value to provide the results
of judgment as relay contact output. (This function is effective when the
meter is equipped with a comparison output unit.) For details on the contact
ratings and other specifications, refer to the section “Output Specifications.”
7.2 Analog Output Function
The FD5000 series of unit meters can output an analog signal for an indicated
value (when the meter is equipped with an analog output unit). There are four
output ranges, 0 to 1 V/0 to 10 V/1 to 5 V/4 to 20 mA, from which a selection
can be made using the condition data. In addition, the analog output of the
FD5000 series allows for arbitrary output scaling. This scaling can be achieved
by setting the indication value for an output of the maximum scale value (20
mA for 4–20 mA output range) in the AOHI parameter of the scaling data.
7.3 RS-485 Interface Function
The FD5000 series can be equipped with an RS-485 interface (when the
meter is provided with an RS-485 unit). For details on the RS-485 function,
see the separate manual on communication functions.
7.4 RS-232C Interface Function
The FD5000 series can be equipped with an RS-232C interface (when the
meter is provided with an RS-232C unit). For details on the RS-232C
function, see the separate manual on communication functions.
8. Specifications and External Dimensions
8.1 Input Specifications
8.1.1DC Voltage Measuring Unit (range 11)
8.1.2DC Voltage Measuring Unit (ranges 12 to 15)
8.1.3DC Current Measuring Unit
Range Measurement
range
Indication
Highest
resolution
Input
impedance
Maximum
permissible
input
Accuracy
11
± 99.99mV
Offset : ± 9999
Full scale : 0 to ± 9999
10µV
About
100M
±100V
± (0.1% of FS)
Input circuit : Single ended type
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Operating system :
conversion
Range
Measurement
range
Indication
Highest
resolution
Input
impedance
Maximum
permissible
input
Accuracy
12
± 999.9mV
Offset : ± 9999
Full scale : 0 to ± 9999
100µV
About
100M
± 100V
± (0.1% of FS)
Input circuit : Single ended type
13
± 9.999V
1mV
About
1M
± 250V
± (0.15% of FS)
± 250V
± 600V
About
10M
About
10M
10mV
100mV
14
15
± 99.99V
± 600V
Operating system :
conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Range
Measurement
range
Indication
Highest
resolution
Input
impedance
Maximum
permissible
input
Accuracy
23
± 9.999mA
1µA
About
10
± 100mA
Offset : ± 9999
Full scale : 0 to ± 9999
± (0.2% of FS)
± (0.3% of FS)
± 500mA
± 3A
About
1
About
0.1
10µA
± 99.99mA
± 999.9mA
100µA
24
25
Input circuit : Single ended type
Operating system :
conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Instruction Manual for FD5000 Series
5/12
Instruction Manual for FD5000 Series
8/12
Condition data
Comparator data
Scaling data
Linearization data
Measurement
Pressing the ENTER key saves the data and returns to the measurement mode.
Calibration data
Shift data
(Data are backed up with EEPROM even when the power is turned off.)
(1) Press the Mode key and the Enter key during
measurement.
(2) Press the Shift key a few times to move to the
linearization data menu.
(4) Press the Shift key (change digit) and the
Increment key (change numeric value) and then press
the Mode Key to after the number of data to be
corrected has been set.
&
Note: The decimal point in the selected digit
flashes.
(3) Press the Mode key to move to the setup for the
number of data to be corrected.
&
(10) When all the settings have been made, press the
Enter key to return to the measurement mode.
(8) Press the Mode key to display the setup for output
values of the data to be corrected.
Note: For a single display unit, the unit
automatically returns to N-01 indication if there is
no key operation for about 8 seconds. In this case,
press the Mode key to return to the numeric value
indication.
&
(5) For a single display unit, press the Mode key to
change to the parameter information indication.
(6) Press the Shift key (change digit) and Increment
key (change numeric value) and then press the Mode
key after the N-01 input value has been set.
RE
Note: For a single display unit, RE flashes when
the input value is set and DZ flashes when the
output value is set.
DZ
&
(7) Press the Shift key (change digit) and Increment
key (change numeric value) and then press the Mode
key after the N-01 output value has been set.
RE
Note: For a single display unit, the unit is
automatically returned to N-02 indication if there is
no key operation for about 8 seconds. In this case,
press the Mode key to return to the numeric value
indication.
(9) Press the Shift key (change digit) and Increment
key (change numeric value) and then press the Mode
key after the N-02 input value has been set.
Repeat steps (5) to (9) until all the settings have
been made.
Multi-display unit
Single display unit
Default Equipped
Indication
Name
value
as
01
02
03
04
05
06
07
08
09
10
11
12
13 14
15 16 17
18
0
1
2
3
4
5
PVH
Peak hold setup
PH
¡
RANG
Measurement range setup
*1
X
¡
¡
¡
¡
¡
¡
¡
X
¡
X
¡
¡
¡
¡ ¡
X
¡
11
15
25
13
15
13
15
25
26
25
26
14
B JPB 14 14
2A
AVG
Number of averaging
1
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
X
X
¡
¡
MAV
Number of moving averaging setup
OFF
¡
S.UD
Step wide setup
1
¡
BLNK
Indication blank setup
OFF
¡
UNIT
Unit setup
C
X
X
X
X
X
X
X
X
X
X
X
X
¡
¡
X
X
X
X
BAUD
Baud rate setup
9600
X
X
X
¡ ¡
X
DATA
Data length setup
7
X
X
X
¡ ¡
X
P.BIT
Parity bit setup
E
X
X
X
¡ ¡
X
S.BIT
Stop bit setup
2
X
X
X
¡ ¡
X
T-
Delimiter setup
CR.LF
X
X
X
¡ ¡
X
ADR
Equipment ID setup
00
X
X
X
X
¡
X
A.OUT
Analog output setup
OFF
X
X
¡
X
X
¡
B.UP
Digital zero backup setup
OFF
¡
LINE
Linearization setup
CLR
¡
I.SEL
Input selection
OC
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡
X
X
X
TR T
Tracking zeroing time setup
00
¡
X
X
TR V
Tracking zeroing width setup *2
01
¡
X
X
SNSR
Sensor power setup
10
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡
X
P
ÇnÇm
Power-on delay setup
OFF
¡
PRO
Protect setup
OFF
¡
U-NO.
Unit number Indication setup
ON
¡
S-HI
HI side judgment value setup
1000
X
¡
X
X
X
¡
S-LO
LO side judgment value setup
500
X
¡
X
X
X
¡
H-HI
HI side hysteresis setup
0
X
¡
X
X
X
¡
H-LO
LO side hysteresis setup
0
X
¡
X
X
X
¡
FSC
Full scale Indication value setup
*1
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
X
X
X
X
X
¡
9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999
9999
FIN
Full scale input value setup
*1
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
X
X
X
X
X
¡
9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999
*3
OFS
Offset indication value setup
*1
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
X
X
X
X
X
¡
0
0
0
0
0
0
0
0
0
0
0
0
0
OIN
Offset input value setup
*1
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
¡
X
X
X
X
X
¡
0
0
0
0
0
0
0
0
0
0
0
0
*4
PS
Pre-scaling value setup
1
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡ ¡
X
X
PPR
Frequency division setup
1
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡ ¡
X
X
DLHI
Digital limiter HI value setup
9999
¡
X
X
X
X
X
X
X
X
DLLO
Digital limiter LO value setup
-9999
¡
X
X
X
X
X
X
X
X
AOHI
Analog output HI indication setup
9999
X
X
¡
X
X
¡
AOLO
Analog output LO indication setup
0
X
X
¡
X
X
¡
DEP
Decimal point position setup
None
¡
*5
¡
ZERO
Zero input value *6
0
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡
X
SPIN
Span input value *6
2000
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡
X
SPAN
Span indication
9000
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X
¡
X
SHF
Shift data setup
0
¡
*1
Each value in the lower part of a cell in the columns on the right is the default value.
*2
Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).
*3
5000 for 1 V range and 2000 for 2 A range
*4
1000 for 1 V range and 400 for 2 A range
*5
Linearization data are not set up for the default values.
*6
This value is not indicated if calibration is done using an actual load.
The shaded parts show the parameters that must be set for each unit.
Shift data
Input unit number
Condition data
Comparator data
Scaling data