background image

5.2 Moving to the Parameter Setup Mode

5.3 Data Lists and Default Settings

5.4.5    Method  of  Setting  Linearization  Data

The linearization function means a function that changes the slope of straight

lines in the relationship between the input and indication by correcting the rela-

tions at arbitrary points.

  

Linearization data are set using the input value (indicated

value before correction) and the output value (indicated value after correction) at
each arbitrary point.

Note: The setup conditions are N-1 < N-2 … N-15 < N-16.

6. Control Functions

6.1 Hold Function

The Hold function temporarily retains the indication.  The hold function is enabled

by shortcircuiting the HOLD and COM terminals or setting both terminals to the same

voltage level.  As a result the display unit retains the indication given at that moment.

6.2 Digital Zero Function

The Digital Zero function zeros the indication given at an arbitrary timing.

Thereafter, the function shows the amount of change from the point of zero-

ing.  However, this function serves as an indication resetting function for a

frequency measuring unit.  Thus, the Digital Zero function can be used to reset

the indication when there is no input signal at all.

Note that, the on/off control of the Digital Zero function can be achieved by

means of terminal control or front panel keys.  In the case of terminal control,

the Digital Zero function is turned on by shortcircuiting the DZ and COM termi-

nals or setting both terminals to the same voltage level.  The indication at that

moment is zeroed.  In the case of control with the front panel keys, hold down

the Mode key and press the Increment key for about 1 second to zero the indi-

cation at that moment.

Note: Operation with the control terminals takes priority over operation

with the front panel keys.  The Digital Zero function is disabled if

the  control  terminals  are  made  to  go  through  the  off-on-off  se-

quence with the function enabled by means of the front panel keys.

6.3 Peak Hold Function

The Peak Hold function retains one of the maximum (peak hold)/minimum

(valley hold)/maximum - minimum (peak-valley hold) values and provides

output for that value.  Selection from these values is made using the condition

data.  The peak hold function is enabled by shortcircuiting the PH and COM
terminals or setting both terminals to the same voltage level.

7.

Output Function

7.1 Comparison Output Function

The FD5000 series of unit meters is designed so that the two judgment values
HI and LO can be set for the measured (indicated) value to provide the results
of judgment as relay contact output.  (This function is effective when the
meter is equipped with a comparison output unit.)  For details on the contact
ratings and other specifications, refer to the section “Output Specifications.”

7.2 Analog Output Function

The FD5000 series of unit meters can output an analog signal for an indicated
value (when the meter is equipped with an analog output unit).  There are four
output ranges, 0 to 1 V/0 to 10 V/1 to 5 V/4 to 20 mA, from which a selection
can be made using the condition data.  In addition, the analog output of the
FD5000 series allows for arbitrary output scaling.  This scaling can be achieved
by setting the indication value for an output of the maximum scale value (20
mA for 4–20 mA output range) in the AOHI parameter of the scaling data.

7.3 RS-485 Interface Function

The  FD5000  series  can  be  equipped  with  an  RS-485  interface  (when  the
meter is provided with an RS-485 unit).  For details on the RS-485 function,
see  the  separate  manual  on  communication  functions.

7.4 RS-232C Interface Function

The FD5000 series can be equipped with an RS-232C interface (when the
meter  is  provided  with  an  RS-232C  unit).    For  details  on  the  RS-232C
function,  see  the  separate  manual  on  communication functions.

8. Specifications and External Dimensions

8.1 Input Specifications

8.1.1DC  Voltage  Measuring  Unit  (range  11)

8.1.2DC  Voltage  Measuring  Unit  (ranges  12  to  15)

8.1.3DC  Current  Measuring  Unit

Range Measurement

range

Indication

Highest

resolution

Input

impedance

Maximum

permissible

input

Accuracy

11

± 99.99mV

Offset : ± 9999

Full scale : 0 to ± 9999

10µV

About

100M

±100V

± (0.1% of FS)

Input circuit : Single ended type
Maximum sampling rate : 12.5 times per second

Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)

Operating system : 

 conversion

Range

Measurement

range

Indication

Highest

resolution

Input

impedance

Maximum

permissible

input

Accuracy

12

± 999.9mV

Offset : ± 9999

Full scale : 0 to ± 9999

100µV

About

100M

± 100V

± (0.1% of FS)

Input circuit : Single ended type

13

± 9.999V

1mV

About

1M

± 250V

± (0.15% of FS)

± 250V

± 600V

About

10M

About

10M

10mV

100mV

14

15

± 99.99V

± 600V

Operating system : 

 conversion

Maximum sampling rate : 12.5 times per second

Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)

Range

Measurement

range

Indication

Highest

resolution

Input

impedance

Maximum

permissible

input

Accuracy

23

± 9.999mA

1µA

About

10

± 100mA

Offset : ± 9999

Full scale : 0 to ± 9999

± (0.2% of FS)

± (0.3% of FS)

± 500mA

± 3A

About

1

About

0.1

10µA

± 99.99mA

± 999.9mA

100µA

24

25

Input circuit : Single ended type

Operating system : 

 conversion

Maximum sampling rate : 12.5 times per second

Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)

Instruction Manual for FD5000 Series

5/12

Instruction Manual for FD5000 Series

8/12

Condition data

Comparator data

Scaling data

Linearization data

Measurement

Pressing the ENTER key saves the data and returns to the measurement mode.

Calibration data

Shift data

(Data are backed up with EEPROM even when the power is turned off.)

(1) Press the Mode key and the Enter key during

measurement.

(2) Press the Shift key a few times to move to the

linearization data menu.

(4) Press the Shift key (change digit) and the

Increment key (change numeric value) and then press

the Mode Key to after the number of data to be

corrected has been set.

&

Note:  The decimal point in the selected digit

flashes.

(3) Press the Mode key to move to the setup for the

number of data to be corrected.

&

(10) When all the settings have been made, press the

Enter key to return to the measurement mode.

(8) Press the Mode key to display the setup for output

values of the data to be corrected.

Note:  For a single display unit, the unit

automatically returns to N-01 indication if there is

no key operation for about 8 seconds.  In this case,

press the Mode key to return to the numeric value

indication.

&

(5) For a single display unit, press the Mode key to

change to the parameter information indication.

(6) Press the Shift key (change digit) and Increment

key (change numeric value) and then press the Mode

key after the N-01 input value has been set.

RE

Note:  For a single display unit, RE flashes when

the input value is set and DZ flashes when the

output value is set.

DZ

&

(7) Press the Shift key (change digit) and Increment

key (change numeric value) and then press the Mode

key after the N-01 output value has been set.

RE

Note:  For a single display unit, the unit is

automatically returned to N-02 indication if there is

no key operation for about 8 seconds.  In this case,

press the Mode key to return to the numeric value

indication.

(9) Press the Shift key (change digit) and Increment

key (change numeric value) and then press the Mode

key after the N-02 input value has been set.

Repeat steps (5) to (9) until all the settings have

been made.

Multi-display unit

Single display unit

Default Equipped

Indication

Name

value

as

01

02

03

04

05

06

07

08

09

10

11

12

13 14

15 16 17

18

0

1

2

3

4

5

PVH

Peak hold setup

PH

¡

RANG

Measurement range setup

*1

X

¡

¡

¡

¡

¡

¡

¡

X

¡

X

¡

¡

¡

¡ ¡

X

¡

11

15

25

13

15

13

15

25

26

25

26

14

B JPB 14 14

2A

AVG

Number of averaging

1

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

X

X

¡

¡

MAV

Number of moving averaging setup

OFF

¡

S.UD

Step wide setup

1

¡

BLNK

Indication blank setup

OFF

¡

UNIT

Unit setup

C

X

X

X

X

X

X

X

X

X

X

X

X

¡

¡

X

X

X

X

BAUD

Baud rate setup

9600

X

X

X

¡ ¡

X

DATA

Data length setup

7

X

X

X

¡ ¡

X

P.BIT

Parity bit setup

E

X

X

X

¡ ¡

X

S.BIT

Stop bit setup

2

X

X

X

¡ ¡

X

T-

Delimiter setup

CR.LF

X

X

X

¡ ¡

X

ADR

Equipment ID setup

00

X

X

X

X

¡

X

A.OUT

Analog output setup

OFF

X

X

¡

X

X

¡

B.UP

Digital zero backup setup

OFF

¡

LINE

Linearization setup

CLR

¡

I.SEL

Input selection

OC

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡

X

X

X

TR T

Tracking zeroing time setup

00

¡

X

X

TR V

Tracking zeroing width setup *2

01

¡

X

X

SNSR

Sensor power setup

10

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡

X

P

ÇnÇm

Power-on delay setup

OFF

¡

PRO

Protect setup

OFF

¡

U-NO.

Unit number Indication setup

ON

¡

S-HI

HI side judgment value setup

1000

X

¡

X

X

X

¡

S-LO

LO side judgment value setup

500

X

¡

X

X

X

¡

H-HI

HI side hysteresis setup

0

X

¡

X

X

X

¡

H-LO

LO side hysteresis setup

0

X

¡

X

X

X

¡

FSC

Full scale Indication value setup

*1

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

X

X

X

X

X

¡

9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999

9999

FIN

Full scale input value setup

*1

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

X

X

X

X

X

¡

9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999

*3

OFS

Offset indication value setup

*1

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

X

X

X

X

X

¡

0

0

0

0

0

0

0

0

0

0

0

0

0

OIN

Offset input value setup

*1

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

¡

X

X

X

X

X

¡

0

0

0

0

0

0

0

0

0

0

0

0

*4

PS

Pre-scaling value setup

1

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡ ¡

X

X

PPR

Frequency division setup

1

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡ ¡

X

X

DLHI

Digital limiter HI value setup

9999

¡

X

X

X

X

X

X

X

X

DLLO

Digital limiter LO value setup

-9999

¡

X

X

X

X

X

X

X

X

AOHI

Analog output HI indication setup

9999

X

X

¡

X

X

¡

AOLO

Analog output LO indication setup

0

X

X

¡

X

X

¡

DEP

Decimal point position setup

None

¡

*5

¡

ZERO

Zero input value *6

0

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡

X

SPIN

Span input value *6

2000

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡

X

SPAN

Span indication

9000

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

X

¡

X

SHF

Shift data setup

0

¡

*1

Each value in the lower part of a cell in the columns on the right is the default value.

*2

Tracking zero width setup parameter is not indicated if the tracking time is set to OFF(0).

*3

5000 for 1 V range and 2000 for 2 A range

*4

1000 for 1 V range and 400 for 2 A range

*5

Linearization data are not set up for the default values.

*6

This value is not indicated if calibration is done using an actual load.

The shaded parts show the parameters that must be set for each unit.

Shift data

Input unit number

Condition data

Comparator data

Scaling data

Содержание FD5000 series

Страница 1: ... V range 11 99 99 mV range 12 999 9 mV 05 AC voltage measuring unit average rms range 14 99 99 V range 15 600 V 06 AC voltage measuring unit true rms range 11 99 99 mV range 12 999 9 mV 07 AC voltage measuring unit true rms range 14 99 99 V range 15 600 V 08 AC current measuring unit average rms range 23 9 999 mA range 24 99 99 mA 09 AC current measuring unit average rms range 23 9 999 mA range 24...

Страница 2: ...inal 1 2 3 4 5 Name Description 1 23 Positive input terminal for range 23 9 999 mA 2 24 3 25 No 4 LO 5 Negative input terminal Positive input terminal for range 24 99 99 mA Positive input terminal for range 25 999 9 mA 5 8 mm min 5 8mm min Applicable solderless terminals 1 2 Name Description No 1 HI Input terminal 2 LO Input terminal 1 2 3 4 5 Name Description 1 HI Input terminal for all ranges 2 ...

Страница 3: ...wire for the power input except for range 26 external control and comparison output connectors 2 Tighten the screws for the power input except for range 26 external control and comparison output connectors to a torque of 0 5 to 0 6 Nm 3 Use 16 to 28 AWG wire for the analog output connector Tighten the screws of analog output connector to a torque of 0 22 to 0 25 Nm 4 Caution Instruction Manual for...

Страница 4: ...ment indicators HI 2 GO LO Indicates the result of judgment and turns on if the measured value HI judgment value Indicates the result of judgment and turns on if LO judgment value the measured value HI judgment value Indicates the result of judgment and turns on if the measured value LO judgment value 3 Function indicators ME Turns on if digital zero backup is on PH Turns on if peak hold valley ho...

Страница 5: ...side judgment value setup 500 X X X X H HI HI side hysteresis setup 0 X X X X H LO LO side hysteresis setup 0 X X X X FSC Full scale Indication value setup 1 X X X X X 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 9999 FIN Full scale input value setup 1 X X X X X 9999 9999 9999 9999 9999 9999 9999 9999 5000 9999 5000 9999 3 OFS Offset indication value setup 1 X X X X X 0 0 0 0 0 0 0 ...

Страница 6: ...g measurement 2 Press the Mode key to change to the peak hold setup mode 3 For a single display unit press the Mode key to change to the parameter information indication The display automatically changes to this indication in about 1 second except right after COND is indicated 4 Press the Increment key a few times to set to Valley Hold 5 Press the Enter key to return to measurement mode Pressing t...

Страница 7: ...ress the Enter or Mode key to return to the measurement mode 5 Press the Shift key change digit and Increment key change numeric value to set 8000 Note The decimal point in the selected digit flashes 3 Press the Mode key while applying pressure that will cause the display to show zero 4 For a single display unit press the Mode key to change to the parameter information indication or 1 Press the Sh...

Страница 8: ...For details on the RS 485 function see the separate manual on communication functions 7 4 RS 232C Interface Function The FD5000 series can be equipped with an RS 232C interface when the meter is provided with an RS 232C unit For details on the RS 232C function see the separate manual on communication functions 8 Specifications and External Dimensions 8 1 Input Specifications 8 1 1DC Voltage Measur...

Страница 9: ...e speed About 1 second Dead zone 0 to 99 digits Crest factor 4 1 at full scale Range Measurement range Indication Highest resolution Input impedance Maximum permissible input Accuracy 23 9 999mA 1µA About 10 100mA Offset 9999 Full scale 0 to 9999 0 5 of FS 20digit 500mA 3A About 1 About 0 1 10µA 99 99mA 999 9mA 100µA 24 25 Input circuit Single ended type Operating system conversion Maximum samplin...

Страница 10: ...o 10s 1s 100ms Renewal time of a display Input type Input voltage lebel Maximum permissible input Open collector Logic Magnet LO 1V or less 5V 2 2 k pull up LO 1V or less HI 2 5 to 15V 0 3 to 30Vp p 30V 15V Duty ratio 50 7 segment LED display character height 14 2 mm on main display and 8 mm on sub display Automatically indicated when the calculated result is negative OL or OL for input signals ou...

Страница 11: ...its ASCII code No procedure Error detection RS 232C RS 485 Start and stop synchronization Full duplex Communication system Communication rate 38400bps 19200bps 9600bps 4800bps 2400bps Start bit 1bit Data length Two wire half duplex Polling selecting system 1 Up to 31 meters log Output 8 3 3 Communicating Function 8 4 External Dimensions 9 Warranty and After service 9 1 Warranty The warranty period...

Страница 12: ... 2002 09 Fuji Electric Instruments Co Ltd International Sales Dept No 1 Fuji machi Hino city Tokyo 191 8502 Japan Phone 81 42 585 6201 6202 Fax 81 42 585 6187 Instruction Manual for FD5000 Series UU 33555a 12 12 ...

Отзывы: