3.3.10 Temperature Measuring Unit (RTD)
3.3.11 Frequency Measuring Unit (Open collector, logic, and magnet)
3.3.12
Frequency Measuring Unit (500 Vrms)
3.3.13
Strain Gauge Input Unit (Load cell)
3.3.14
Process Signal Measuring Unit
3.4 Comparison Output
Connection of three-wire sensor
1 2 3
Name
Description
1
A
Resistance sensor wire
2
B
3
No.
C
Elimination of wire resistance
Resistance sensor wire
1 2 3
Name
Description
1
HI
Input terminal
2
NC
3
No.
LO
Do not connect this terminal.
Input terminal
1 2 3 4 5
Name
Description
1
+SIG Positive input terminal
2
-SIG
3
+EXC
No.
4
COM
5
Common terminal (grounding terminal for
input circuit)
Negative input terminal
Power output for sensor (positive)
-EXC Power output for sensor (negative)
1 2 3
Name
Description
1
V-IN Positive input terminal for 1 to 5 V range
2
A-IN
3
No.
LO
Positive input terminal for 4 to 20 mA range
Negative input terminal
12 13 14 15 16 17 18 19
Name
Description
12
LO-b LO output terminal (b contact)
13
LO-c
14
No.
LO-a
Common terminal for LO output
LO output terminal (a contact)
15
GO-c Common terminal for GO output
16
GO-a
17
HI-b
GO output terminal (a contact)
HI output terminal (b contact)
18
HI-c
19
HI-a
Common terminal for HI output
HI output terminal (a contact)
3.5 Analog Output
3.6 Serial Communication
4. Components and their Functions
The front panel design of the FD5000 series of unit meters differs depend-
ing on the display unit selected. The names and functions of each unit are
as shown below.
4.1 Multi-display Unit
24 2526
Name
Description
24
COM Common terminal for analog output
25 A-OUT
26
No.
V-OUT
Current output terminal (4 to 20 mA)
Voltage output terminal (1 to 5 V, 0 to 1 V,
and 0 to 10 V)
20212223
Name
Description
20 RXD(+) RS-232C: transmission;
RS-485: Non-reverse output
21
TXD(-)
22
No.
NC
RS-232C: reception;
RS-485: Reverse output
Do not connect this terminal.
23
SG
Common terminal for communications
(1) Use 12 to 28 AWG wire for the power, input (except for range 26),
external control, and comparison output connectors.
(2) Tighten the screws for the power, input (except for range 26),
external control, and comparison output connectors to a torque of
0.5 to 0.6 Nm.
(3) Use 16 to 28 AWG wire for the analog output connector.
Tighten the screws of analog output connector to a torque of 0.22
to 0.25 Nm.
(4)
Caution
!
Range
Input
sensor
Indication
Highest
resolution
Accuracy
PA
0.1°C
(0.1°F)
± (0.15% of FS)
Input circuit : Single ended type
JPA
PB
JPB
Operating system :
" #
conversion
Maximum sampling rate : 12.5 times per second
PT100
!
JPt100
!
-100.0 to 199.9°C
(-148.0 to 391.8°F)
-100 to 600°C
(-148 to 1112°F)
1°C
(1°F)
± (0.3% of FS)
JPt100
!
PT100
!
Current through RTD : About 1 mA
External resistance : 10
!
or less per wire
Linearizer : Digital linearizer
Burnout alarm : It blinks by ---- display.
Range
Indication
Highest
resolution
Accuracy
11
12
13
14
0.1 to 200Hz
1 to 2000Hz
Pre-scale : 0.001 to 5
Frequency division : 1 to 100
0.1Hz
± (0.2% of FS)
0.01 to 20kHz
Measurement
range
0.1 to 200kHz
1Hz
10Hz
100Hz
0 to 10s
1s
100ms
Renewal
time of a
display
Input type
Input voltage lebel
Maximum
permissible input
Voltage
500Vrms
Duty ratio : 50%
50 to 500Vrms
Sensor
power
Highest
resolution
Accuracy
5V
10V
-0.3 to +2mV/V
0.5µV/digit
± (0.1% of FS +2digit)
Zero adjusting rangeSpan adjusting range
1 to 3mV/V
1µV/digit
Input circuit : Single ended type
Operating system :
"
#
conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Range Measurement
range
Indication
Input
impedance
Maximum
permissible
input
Accuracy
1V
1 to 5V
Offset : ± 9999
Full scale : 0 to ± 9999
About
100M
!
± 100V
± (0.2% of FS)
2A
4 to 20mA
About
10
!
± 100mA
Input circuit : Single ended type
Operating system :
"
#
conversion
Maximum sampling rate : 12.5 times per second
Noise rejection ratio : NMR (normal mode rejection) 50 dB or more (50 or 60 Hz)
Instruction Manual for FD5000 Series
3/12
Instruction Manual for FD5000 Series
10/12
8.1.14
Temperature Measuring Unit (RTD)
8.1.15 Frequency Measuring Unit (open collector, logic, and magnet)
8.1.16 Frequency Measuring Unit (500 Vrms)
8.1.17 Strain Gage Unit
8.1.18 Process Signal Measuring Unit
8.2
Common Specifications
8.3
Output Specifications
8.3.1 Output for Comparison
Range
Indication
Highest
resolution
Accuracy
11
12
13
14
0.1 to 200Hz
1 to 2000Hz
Pre-scale : 0.001 to 5
Frequency division : 1 to 100
0.1Hz
±(0.2% of FS)
0.01 to 20kHz
Measurement
range
0.1 to 200kHz
1Hz
10Hz
100Hz
0 to 10s
1s
100ms
Renewal
time of a
display
Input type
Input voltage lebel
Maximum
permissible input
Open collector
Logic
Magnet
LO : 1V or less (5V : 2.2k
!
pull up)
LO : 1V or less,HI : 2.5 to 15V
0.3 to 30Vp-p
30V
15V
Duty ratio : 50%
7-segment LED display (character height : 14.2 mm on main
display and 8 mm on sub-display)
Automatically indicated when the calculated result is
negative.
OL or -OL for input signals outside the indication range
-9999 to 9999
Leading zero suppression
HOLD, PH, DZ (reset for frequency measuring unit)
-10 to 70 °C, 60% RH or less
100 to 240 V AC ± 10% for AC power supply unit
9 to 50 V DC for DC power supply unit
Approx. 5 W
96 mm (W) x 48 mm (H) x 146.5 mm (D)
Note : Depth (D) denotes the maximum value.
Weight :
2000 V AC for 1 min. between power terminals and
input terminal, and between power terminals and each
output terminal (AC power supply)
500 V DC for 1 min. between power terminals and input
terminal, and between power terminals and each output
terminal (DC power supply)
100 M
!
between the above terminals when 500 V DC
is applied
500 V DC for 1 min. between input terminal and each
output terminal, and between analog output terminal and
communication terminals
2000 VAC for 1 min between case and each output
terminal (common to both AC and DC supply)
Display :
Polarity indication :
Indication range :
Over-range alarm :
Decimal point : Can be set at an arbitrary digit.
Zero indication :
External control :
Operating temperature and humidity range : 0 to 50°C, 35 to 83% RH
(non-condensing)
Storage temperature and humidity range :
Power supply :
Power consumption :
External dimensions :
450g
Withstand voltage :
Withstand voltage
:
:
Withstand voltage
Insulation
resistance
:
Conditions for comparison
Indicated value > Upper limit judgment value
Judgment
result
HI
Lower limit judgment value
$
Indicated value
$
Upper limit judgment value
GO
Lower limit judgment value > Indicated value
LO
Control system : Micro computer operating system
Judgment value setup range :-9999 to +9999
Hysteresis : Can be set in the range of 1 to 999 digits for each judgment
value
Operating speed : Depends on the sampling rate.
Output method : Relay contact output (Make and break contacts for HI and
LO and make contacts for GO)
Output rating : 240 V AC, 8 A (resistive load) and 30 V DC, 8 A (resistive
load)
1 2 3 4 5
Name
Description
1
HI
Positive input terminal
2
LO
3
+15V
No.
4
COM
5
Common terminal (grounding terminal for
input circuit)
Negative input terminal
Power output for sensor (positive)
0V
Power output for sensor (negative)
(1)
(2)
(3)
(4)
(6)
(7)
(8)
(9)
(5)
No.
Name
Main Functions
During measurement
During parameter setup
(1) Main display
Indicates the measured value.
IIndicates information on the
parameter to be set.
Judgment
indicators
HI
(2)
GO
LO
Indicates the result of judgment and
turns on if the measured value > HI
judgment value.
Indicates the result of judgment and
turns on if LO judgment value the
measured value HI judgment value.
Indicates the result of judgment and
turns on if the measured value < LO
judgment value.
(3) Function
indicators
ME
Turns on if "digital zero backup" is on.
PH
Turns on if "peak hold/valley hold/peak -
valley hold" is on.
Turns on if "digital zero" is on.
DZ
RE
Turns on if remote control is being
performed through RS-232C or RS-485
interface.
(4) Sub-display 1
Indicates the HI side judgment value.
Indicates the item in the maximum/
minimum/(maximum-minimum)/input
value monitoring mode.
Indicates the LO side judgment value.
Sub-display 2
(5)
Indicates information on the item in the
maximum/minimum/(maximum-
minimum)/input value monitoring mode.
Indicates the item to be set.
(6) Enter key
Pressing the Enter and Mode keys
together changes to the parameter
setting mode.
Pressing the Enter and Increment keys
together changes to the maximum/
minimum/(maximum-minimum)/input
value monitoring mode.
Switches from the maximum/minimum/
(maximum-maximum/minimum/
(maximum-minimum)/input value
monitoring mode to the comparative
judgment reading mode.
Returns to the measurement
mode.