Isolation Systems Inc
Operation and Maintenance Manual PASS-1600ISO
2 October, 2020
Isolation Systems Inc
13
889 Erie Ave, North Tonwanda NY, 14120
Intake
Controller
Exhaust
Set-up
Large Intake: Patient Room, No Duct
Small Intake: Ducted to Ante-room
Large Exhaust: Patient Room, No Duct
Small Exhaust: Ducted outside
Dampers: Intake 1/3 open, Exhaust closed
Controller: Pressure tube outside shelter
Unit Inside Shelter
Positive Pressure
Room
Exhaust
Intake
Controller
Unit Outside Shelter
Positive Pressure
Set-up
Large Intake: Ducted to Patient Room Small
Intake: Ducted to Ante-room
Large Exhaust: Ducted to Patient Room
Small Exhaust: No Duct
Dampers: Intake 1/3 open, Exhaust closed
Controller: Pressure tube outside shelter
Room
Intake
Controller
Exhaust
Unit Inside Shelter
Air Filtration/Circulation
Set-up
Large Intake: Patient Room, No Duct
Small Intake: Ducted to Ante-room
Large Exhaust: Patient Room, No Duct
Small Exhaust: Ducted outside
Dampers: Closed
Controller: Alarm Disabled
Room