Isolation Systems Inc
Operation and Maintenance Manual PASS-1600ISO
2 October, 2020
Isolation Systems Inc
12
889 Erie Ave, North Tonwanda NY, 14120
Exhaust
Intake
Ante-room
Controller
Unit Outside Shelter
Negative Pressure with Ante-room
Set-up
Large Intake: Ducted to Patient Room
Small Intake: Ducted to Ante-room
Large Exhaust: Ducted to Patient Room
Small Exhaust: No Duct
Dampers: Exhaust 1/3 open, Intake closed
Controller: Pressure tube outside shelter
Room
Intake
Controller
Exhaust
Unit Inside Shelter
Negative Pressure
Set-up
Large Intake: Patient Room, No Duct
Small Intake: Ducted to Ante-room
Large Exhaust: Patient Room, No Duct
Small Exhaust: Ducted outside
Dampers: Exhaust 1/3 open, Intake closed
Controller: Pressure tube outside shelter
Room
Exhaust
Intake
Controller
Unit Outside Shelter
Negative Pressure
Set-up
Large Intake: Ducted to Patient Room Small
Intake: Ducted to Ante-room
Large Exhaust: Ducted to Patient Room
Small Exhaust: No Duct
Dampers: Exhaust 1/3 open, Intake closed
Controller: Pressure tube outside shelter
Room