3.0 Disposal / Sanitizing / Decontamination
Standard Procedures
3.1
Handle any contaminated materials in accordance with your institution’s or agency’s
regulations.
3.2
If the Unit or any materials can be sanitized or decontaminated, follow your institution’s or
agency’s regulations and procedures.
3.3
Dispose of all contaminated materials or equipment in accordance with your institution’s or
agency’s regulations and procedures. Dispose of HEPA filters when the filter pressure is 1.6
in WC, or no less often than 8760 hours of run time or annually as is provided with your
service contract.
4.0 Maintenance (Scheduled and Unscheduled)
WARNING!
Only a trained, authorized maintenance person should access the filter
compartment. Only a trained, authorized technician should access elsewhere
.
Tools required:
•
Standard screwdriver for access to filter cabinet
•
5/16 nut driver for access to electrical cabinet.
•
3/16-inch Allen wrench
Before any maintenance procedure, disconnect all power from the unit. Power to the unit will
result in dangerous electrical contact and exposure to dangerous UV radiation. For any
service requiring power to the UV lamps, you must use safety glasses rated for UV protection
and ensure that no one without UV rated safety glasses can be exposed to the radiation.
Access to the filter compartment may expose biological
contaminated waste materials. Wear PPE as required and
be prepared to “red bag” waste that is removed.