INDICATIONS FOR USE& GENERAL SAFETY  

 

1.1 

Indications for Use 

 

 

The Medical Treatment Chair, T688B Series chair is intended for use in medical procedures such 

as  the  administration  of  renal  dialysis  to,  and  collecting  blood  from,  patients  in  hospital 
departments, under the supervision of trained medical staff. The Medical Treatment Chair is also 
intended  for  use  in  day  surgery.  The  T688B  Series  chair  is  designed  so  that  the  occupant  is 
accommodated  in  a  seated  position  with  the  hips  moved  back  so  that  the  occupant’s  back  is 
against  the  back  rest  and  the  legs  outstretched  and  supported  by  the  seat  and  leg  rests.  The 
chair is designed in such a way that no special training nor competence is needed for operating it 

The T688B Series chair is also used to position patients for easy access by healthcare 
professionals. 

The T688B Series chair is intended to be used by patients with a weight not exceeding 300kg 
(661 lbs).  

 

1.2 

General Safety 

 

The  T688B  Series  chair  is  intended  to  be  used  in  Professional  Healthcare  Environments 
(e.g. hospitals). 

The T688B Series chair shall only be used indoors on flat surfaces. 

The patient should, at all times, be under the supervision of a healthcare professionals.   

The chair should never be used in a manner for which it was not intended. 

The chair is designed so that the occupant will be seated in a typical seated position - hips moved 
back so that the occupant’s spine is against the backrest, with legs outstretched and supported 
by the seat and leg rest.  

The  chair  is  fitted  with  castors,  which  are  specifically  used  to  aid  cleaning  and/or  positioning  of 
the chair within the room.  

The T688B Series chair shall not be used to transport patients or 

any other items

.  

Ensure that the backrest is fully elevated before moving the chair. 

For models fitted with dual hand controls: the programmable unit (

Nurse’s Hand Control) is for the 

use of the staff members only, and the dual recline unit (

Patient’s Hand Control), where fitted, is 

to  be  used  by  the  patient.   

This  ensures  that  the  chair  is  not  raised  or  lowered,  by  the 

occupant.  A  staff  member  should  ensure  the  area  surrounding  the  chair  is  free  from 
obstruction  before  adjusting  it. 

The Nurse’s Hand Control should be kept out of reach of 

the patient at the rear of the chair. 

It is the staff  member's responsibility to  ensure that prior to operating  the chair,  the surrounding 
walls or equipment will not come into contact with nor obstruct the free movement of the chair. 

It is the staff members' responsibility to ensure the patient is briefed on the safe operation 
of  the  chair.    If  the  staff member  believes  the  patient  cannot  safely  operate  the  chair,  the 
staff member should remove the h

and control from the patient’s reach.  

In  an  emergency,  the  chair  can  be  lowered  to  the  Trendelenburg  position  where  CPR  may  be 
performed. It is the staff members' responsibility to ensure that the chair is in the correct position 
for CPR and that the resuscitation method used 

is in line with hospital policy.

  In addition, staff 

should  ensure  that  the  chair  is  adequate  for  those  resuscitation  methods, 

in  line  with  hospital 

policy. Do not sit on the backrest or the leg rest of the chair. 

The upholstery is applied parts. The upholstery has no electrical connection to other parts of the 

Содержание T688B Series

Страница 1: ...Medical Treatment Chair T688B Series Instruction Manual ...

Страница 2: ...ND CONTROLS PROGRAMMING 7 3 3 1 Hand Control 7 3 3 2 Programming Instructions 8 3 3 3 Recalibration 8 3 MAINTENANCE 9 4 1 INSPECTION SERVICING 9 4 2 CLEANING 9 4 3 FABRIC STANDARDS 10 4 4 TROUBLE SHOOTING 10 4 ARRANGING A SERVICE 11 5 WARRANTY INFORMATION 12 6 WASTE DISPOSAL 12 7 PHYSICAL DESCRIPTION 13 8 1 DESCRIPTION OF DEVICE 13 8 2 TECHNICAL SPECIFICATIONS 13 8 2 1 Dimensions Weight Capacity 1...

Страница 3: ...the seat and leg rest The chair is fitted with castors which are specifically used to aid cleaning and or positioning of the chair within the room The T688B Series chair shall not be used to transport patients or any other items Ensure that the backrest is fully elevated before moving the chair For models fitted with dual hand controls the programmable unit Nurse s Hand Control is for the use of t...

Страница 4: ...ntraindication for medical treatment chairs But limitation of use in patients with history of Allergy to covering material of chairs Pressure sores 1 5 Reporting serious incident Any serious incident that has occurred in relation to this medical device shall be reported to the manufacturer and local health authority in accordance with the local regulations Within the EU the user must report any se...

Страница 5: ...as per instruction 5 above 9 Arms maybe swung out or lowered flat depending on style to seat level to facilitate side transfer of patient Raise or lower the seat height to prevent unnecessary bending or lifting by staff members 10 IV Pole if fitted can be removed or repositioned by loosening the knob at the lower end of the mounting bracket and retightening the knob 11 Lift up tray if fitted can b...

Страница 6: ...mergency stop press any button on either the nurse hand control or the patient hand control if fitted 20 USB charger if fitted can be used to charge a USB device such as a mobile phone a music player or a camera Maximum output of the charger is 4 watt NOTE It may not charge all USB devices depending on the power input requirement of the USB devices NOTE To use the USB charger ensure the chair is p...

Страница 7: ...witched on Refer to section 3 2 The backup batteries if fitted will recharge automatically and slow to trickle charge when full It is important to charge the backup battery for a minimum of 24 hours before use the chair 3 1 WARNINGS AND PRECAUTIONS The following table shows potential hazards that have been identified and the steps that should be taken to avoid them IDENTIFIED HAZARD CAUSE PREVENTA...

Страница 8: ...please contact Fresenius Medical Care or your local representative at the details provided in section 5 The battery is not user replaceable The chair must be connected to the power supply to recharge at least every 5 days 3 3 HAND CONTROLS PROGRAMMING Each T688B Series Medical Treatment Chair can be supplied with two hand controls One nurse operated hand control and one patient operated hand contr...

Страница 9: ...calibrateT688B Series chair as follows using the nurse s hand control 1 Press both up and down arrows of the Height function simultaneously and hold until the beeping stops approximately five seconds NOTE both UP and Down buttons must be activated at exactly the same time 2 Press both up and down arrows of the Backrest control function simultaneously and hold until all the actuators have stopped m...

Страница 10: ...material or a sample provided by the manufacturer The following should be used as a guideline only and no claims will be accepted for damage as a result of following these guidelines RECOMMENDED CLEANING PROCEDURE FOR MEDICAL TREATMENT CHAIRS The following recommended procedure should only be used if it generally conforms to Infection Control guidelines of the hospital Step 1 Bodily fluids may car...

Страница 11: ...n correctly as indicated by the green indicator light on the nurse s hand control 3 Ensure that the batteries have been on charge for 24 hours minimum This should occur at least every 5 days 4 Ensure there are no nicks or cuts in the hand control or actuator cables 5 Determine what is not operating as follows The leg motor vertically mounted under the seat Depress the hand control button for the l...

Страница 12: ...ote Please quote the chair serial number which can be found on the Type Label attached to the chair An example of the Type Label is in section 8 2 7 CONTACT INFORMATION https fmc seating com MANUFACTURER SERVICE CENTRE FRESENIUS MEDICAL CARE SEATING AUSTRALIA PTY LTD 786 Stud Road Scoresby VICTORIA 3179 AUSTRALIA sales seatingAU fmc asia com SERVICE TEL 61 0 3 9780 9509 SALES TEL 61 0 3 9780 9500 ...

Страница 13: ...onth warranty 6 The T688B Series Medical Treatment Chairs are not covered by warranty if used outdoors 7 The use of unauthorised labour will void the warranty Contact Fresenius Medical Care for approval prior to commencing work 8 If you are unsure about any aspect of this warranty please contact Fresenius Medical Care at the number given in Section 5 of this manual If you need to arrange a service...

Страница 14: ...g castors or central locking castors removable upholstery of adjustable head rest backrest seat cushions and leg rest metal frame 8 2 TECHNICAL SPECIFICATIONS 8 2 1 Dimensions Weight Capacity T688B Dimensions depending on options Height 128 135 cm Width 112 cm Length 100 cm Weight depending on options 128kg 282 lbs Capacity 300kg 661 lbs 8 2 2 Electrical Safety CB 6 Control Box Specifications Degr...

Страница 15: ...V2 Apr 2021 8 2 4 Operating Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 5Transport Storage Conditions Temperature range 5ºC 40ºC 41ºF 104ºF Relative humidity 20 90 8 2 6 Electromagnetic or other interference Switch off devices which can cause electromagnetic interference ...

Страница 16: ...rticle Number Catalog Number Date of manufacture Serial Number CE mark Manufacturer For indoor use only Caution warning Consult Instruction Manual authorized representative in the European Community Medical Device Degree of protection against electrical shock Type B Storage and operating temperature 5ºC 40ºC Storage and operating relative humidity 20 90 ...

Страница 17: ...tery is in normal use and requires charging LED Amber ON and LED Green ON indicates mains is on and battery is charging LED Amber Off and LED Green Off indicates Mains is not on and battery requires charging Chair will not operate LED Amber blinking and LED Green blinking and indicates a system error Note When the chair is not connected to the mains press any button on the nurse hand control to wa...

Страница 18: ...ns test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The T688B Series Chair uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any Interference In nearby electronic equipment RF emissions CISPR 11 Class B The T688B Series Chair is suitable for use in all RF emissions establishments other than domestic a...

Страница 19: ...wer quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT...

Страница 20: ...e in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propa...

Страница 21: ... to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 3 691 1 17 1 17 2 33 1011 67 3 69 3 69 7 38 100 11 67 11 67 23 33 For transmitters rated at a maximum output power not listed above the recomme...

Страница 22: ...Fresenius Medical Care Seating Australia Pty Ltd 786 Stud Road Scoresby VIC 3179 Australia IFU T688B EN V2 April 2021 ...

Отзывы: