FRESENIUS MEDICAL CARE
MEDICAL TREATMENT CHAIR
INSTRUCTION MANUAL - T688B Series
Page 17 of 20
IFU T688B V1 May 2019
8
Guidance and manufacturer's declaration
Guidance and manufacturer's declaration -electromagnetic emission
The
T688B Series Chair
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
T688B Series Chair
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
T688B Series Chair
uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any Interference In nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The
T688B Series Chair
is suitable for use in all
RF emissions establishments other than domestic and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic
Emissions
IEC 61000-3-2
class A
Voltage
Fluctuations/
flicker emissions
IEC 61000-3-3
Complies