Unlocking the keypad is required to stop feeding, change feeding settings and enter the
menu.
4.2.10 Mute alarm
To temporarily release alarm sound, press
.
When a medium priority alarm is muted:
■
the mute icon
is displayed in the status bar;
■
the alarm symbol is displayed and the yellow LED keeps flashing until a corrective action
is performed;
■
the alarm sound is off for 2 minutes.
When a low priority alarm is muted:
■
the mute icon
is displayed in the status bar;
■
the alarm symbol is displayed and the yellow LED is lit;
■
the alarm sound is off and an information signal sound (2 beeps) is emitted every 30
minutes.
For further information about alarms, see
on page 41.
26