9.
Make sure to use the original door when replacing it on the pump (check the serial number
on the pump is the same as on the door).
6.4.2 Disinfection instructions
Prerequisites
■
The cleaning protocol has been performed.
■
The pump is switched off.
■
The power cord and all other cables are unplugged.
■
The pump is disconnected from the holder.
■
The air is at room temperature (20 to 25 °C).
■
The operator is wearing suitable protective equipment.
Protocol
1.
Place the previously cleaned pump and holder on a cleaned surface or disposable
underlay. The door can be removed from the pump to facilitate the disinfection.
2.
During disinfection, do not turn the pump over to avoid liquid leak in the battery door.
3.
Use a ready-to-use wipe to wipe down all exposed surfaces of the pump, holder and
pump door (as described in cleaning protocol), making sure to cover all cracks, crevices,
and hard-to-reach areas. Do not allow liquids to run, leak, or drip into the pump housing.
4.
Using a fresh ready-to-use wipe, repeat steps 3. Ensure that the minimum contact time
for each step is 3 minutes for bactericide activity (surface remain visibly wet for 3 minutes).
Respect the indicated contact time from the manufacturer recommendations for the
required antimicrobial activity.
5.
Wipe down the power cord.
6.
Allow the pump to dry completely at room temperature.
7.
Make sure to use the original door when replacing it on the pump (check the serial number
on the pump is the same as on the door).
38