Actions
Meanings
Symbol
■
Close pump door.
▷
on page
17.
Pump door not properly closed at start.
Door open
■
Close pump door.
▷
on page
17.
Pump door opened after start.
■
Re-hang door.
Pump door removed from its anchor-
ing.
■
Contact your biomedical depart-
ment.
Door mechanism is faulty.
■
Open the door, check set installa-
tion.
▷
on page
17.
■
Check that the set is not kinked.
■
Check that upstream clamp is
open.
■
Flush tube if necessary.
■
Check the absence of upstream /
downstream occlusion in the line.
Upstream flowpath is blocked between
the container and the pump.
Upstream occlu-
sion
■
Open the door, check the set in-
stallation, close the door.
▷
on page
17.
■
Check that the set is not kinked.
■
Re-position and verify that food
flows freely after adjustment.
■
Check that the feeding tube is
clear.
■
Flush tube if necessary.
■
Check the absence of upstream /
downstream occlusion in the line.
Downstream flowpath is blocked after
the pump, at the patient side.
Downstream occlu-
sion
Feeding control
Actions
Meanings
Symbol
Low priority - Yellow LEDs are fixed
43