4.2 Basic operations
Before using the pump, proceed to Quick Check Protocol (see
on
page 39).
4.2.1 Switch-on
When using a pump on patient requiring special attention, ensure that a backup pump or
gravity set are available for immediate use.
When switching on the pump, check that the auto test sequence is as described below.
Before switching on the pump, install holder and pump, see
on page 12.
Figure 4:
Auto test
During the 2-second autotest:
■
red, yellow and green LEDs blink;
■
beep sounds (if sound level is low, melody is playing on low, if sound level is high, melody
is on high).
4.2.2 Installing the giving set
4.2.2.1 Preparing the giving set
In order to protect user health, please follow clean aseptic handling procedures for container,
set or feeding tube disposal.
WARNING
■
Only Fresenius Kabi giving sets can guarantee pump reliability. Please
refer to the compatible giving sets (see
on page 63) and
compatible nutrition fluids (see
on page 5).
■
Check the giving set's intended use regarding the feeding protocol,
especially for patients requiring special attention.
■
Check giving set and patient connection integrity before use.
17