10
®
D
When used alone the command ‘d’ returns the displayed value. If a number
follows the command from 1 to 10, the returned value will be as follows:
d1 = V
d2 =
d3 = VA
d4 =
f
d5 = Hz
d6 = W
d7 = Ws
d8 = Wh
d9 = time (s)
d10 = time (h)
If the value for
f
(phase angle) cannot be calculated either because the voltage or the
current is too low, the value 0 is returned.
G
Read set-up line. The format is 28 char CR.
The set-up line comprises the following information:
Character no.
Significance
1
Start conditions:
‘1’ = manual start
‘2’ = clock
‘3’ = power level
2-13
Start-time
Format:
ssmmhhddmmyy
ss = sec., mm = minute, hh = hour,
dd = day, mm = month, yy = year
14-19
Power level:
Format: xxxx.xx in Watt
20
Log layout:
‘1’ = automatic compression
’2’ = fixed time
21-28
Logging time:
Format: ssmmhhdd
ss = sec., mm = minute, hh = hour,
dd = day
H
Write set-up line. The set-up line must be entered immediately after the command.
L
Set datalog. On request type the new set-up data. The start conditions must be set
prior to selecting the log layout. Eventually decide whether to store the new set-up
or not.
M
Messages turned on (default off). The Watt & Energy Meter will return a message
when:
The logging function is changed manually.
The log is compressed, turned on or off.
N
Messages turned off (default)
O
one line, read all data