101364
Model
FT2
Use the
CHG (F1)
key to change the entry,
PRV (F2)
to move to the previous entry,
NXT (F3)
to
move to the next entry and
EXIT (F4)
to return.
The calibration table can hold up to 20 data pair points. Each data point has a voltage and mass
velocity associated with it.
Pressing the
NXT (F3)
key will show the data point voltage and then mass velocity and then go to
the next data point. The number after Volt (ie.Volt1) or Flo (i.e. Flo1) indicated the data point
number.
Upon exiting the menu, the display will prompt for a calibration date as follows:
CalDate=2/24/03
CHG OK
F1
F2
F3
F4
4.1.12.
Reset Total and Elapsed Time
The resetting of the totalizer and elapsed time is accomplished by pressing the
F3
and
F4
keys at
the same time in the normal running mode.
RESET TOTAL ?
NO YES
F1
F2
F3
F4
Press
YES (F4)
to reset total and
NO (F1)
to cancel.
If you entered the programming menu with a
Level 2
password,
The following menu
Restore Database
will be displayed:
4.1.13.
Restore Database
Restoring the original factory settings is accomplished from the “Flow Parameter 2” menu
by entering a
Level 2
password “9111” and pressing the
SPC
key (
F2
).
RESTORE DATABASE ?
YES NO
F1
F2
F3
F4
Press
YES (F1)
ONLY
if you want to restore your database to the initial factory setting that the
meter was shipped with. All current user-entered settings will be overwritten.
4.1.14.
Reset CRC
If the EEPROM CRC check fails (Error Code 36), the EEPROM values will need to be verified
and corrected before clearing the error. Call FOX Customer Service if you need assistance.