101364
Model
FT2
After pressing
OK
(F4)
to accept the reference temperature, the display will prompt for the
pressure unit selection:
PRES UNT= mmHG
NXT OK
F1
F2
F3
F4
Press
NXT (F1)
to select next entry and OK (F4) to accept it.
Selections
are:
mmHG
Psia
bara
After the pressure unit selection is made, the display will show a menu to enter the pressure
reference:
PresRef= 760
CHG OK
F1
F2
F3
F4
Press
CHG (F1)
to change it and
OK (F4)
to accept.
After the pressure reference is accepted, the display will prompt for the gas density if LBS or KG
was selected for flow unit:
DNS = 0.988876 KG/m3
CHG OK
F1
F2
F3
F4
Press
CHG (F1)
to change and
OK (F4)
to accept.
Note: The density entry is only used when KG/time or LBS/time is selected for flow rate units.
Density conditions are referenced to 0 C° at 760 mmHg.
4.1.10.
Flow Parameters
This is the menu used to set various flow parameter values. They are:
Flow cutoff, pipe area, filter, high and low alarm for flow and temperature.
The menu is accessed from the base programming menu by pressing
FLO (F2)
key:
SET PARAMETERS
I/O FLO DSP EXIT
F1
F2
F3
F4