5:66
Service Manual 1001 5015 / Rev. 4
5.5.2
Replacement of STM Cuvette Cam Thread
This instruction is valid for STM cuvette part no:
•
10012592 (type 1)
•
10013213 (type 1)
•
10014163 (type 1)
•
10011075 (type 2)
Replacement - type 1
1.
Remove the two grub screws
1
, see Fig. 5:67.
Fig. 5:67 STM cuvette
2.
Remove the old cam thread
1
, see Fig. 5:68, by sliding it down and out through
the open end. If the cam thread is stuck use a screwdriver and a hammer to gently
knock it loose.
Fig. 5:68 Removing cam thread
Содержание Infratec 1241
Страница 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Страница 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Страница 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Страница 75: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 4 7 Fig 4 2 Troubleshooting the I2C communication ...
Страница 76: ...Infratec 1241 Grain Analyzer 4 8 Service Manual 1001 5015 Rev 4 Fig 4 3 I2C Bus in Infratec 1241 ...
Страница 130: ...Infratec 1241 Grain Analyzer 5 2 Service Manual 1001 5015 Rev 4 Fig 5 1 New Display Cable 1 Ferrites ...
Страница 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Страница 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Страница 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Страница 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Страница 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Страница 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...