![Foss Infratec 1241 Скачать руководство пользователя страница 192](http://html1.mh-extra.com/html/foss/infratec-1241/infratec-1241_service-manual_2323170192.webp)
5:64
Service Manual 1001 5015 / Rev. 4
3.
If the level sensor is in a wrong position - loosen the screws
1
. Adjust the level
sensor to correct position and tighten the screws
1
.
Fig. 5:66 Adjusting level sensor
4.
Assemble in reverse order.
5.
Start up the instrument and let it go through the Startup Test.
6.
Perform a TWM balance calibration, see 5.7.2 TWM Balance Calibration on
page 5:74.
1
Screws
1
Содержание Infratec 1241
Страница 8: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 ...
Страница 34: ...Infratec 1241 Grain Analyzer 2 24 Service Manual 1001 5015 Rev 4 ...
Страница 68: ...Infratec 1241 Grain Analyzer 3 34 Service Manual 1001 5015 Rev 4 Fig 3 66 ...
Страница 75: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 4 7 Fig 4 2 Troubleshooting the I2C communication ...
Страница 76: ...Infratec 1241 Grain Analyzer 4 8 Service Manual 1001 5015 Rev 4 Fig 4 3 I2C Bus in Infratec 1241 ...
Страница 130: ...Infratec 1241 Grain Analyzer 5 2 Service Manual 1001 5015 Rev 4 Fig 5 1 New Display Cable 1 Ferrites ...
Страница 203: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 5 75 Fig 5 77 Fig 5 78 Fig 5 79 ...
Страница 206: ...Infratec 1241 Grain Analyzer 5 78 Service Manual 1001 5015 Rev 4 ...
Страница 208: ...Infratec 1241 Grain Analyzer 6 2 Service Manual 1001 5015 Rev 4 ...
Страница 209: ...Infratec 1241 Grain Analyzer Service Manual 1001 5015 Rev 4 7 1 7 Schematics N A ...
Страница 210: ...Infratec 1241 Grain Analyzer 7 2 Service Manual 1001 5015 Rev 4 ...
Страница 212: ...Infratec 1241 Grain Analyzer 8 2 Service Manual 1001 5015 Rev 4 ...