5:4
User Manual 6007 2981 / Rev. 10
5.1.4
Cleaning of Anti-Static Brush
Upon cleaning the belt brush or replacement of the belt, the anti-static brush should
also be cleaned in order to remove excess material that accumulates in it. This is not
prompted by the Clean procedure.
1.
Remove the conveyor belt cover.
2.
The drawer can be removed for easier access to the antistatic brush.
3.
Remove any kernels or parts of kernels stuck to the antistatic brush.
4.
Reassemble conveyor belt cover.
Fig. 5:6
Note: Replacement of anti-static brush is part of the Preventive
Maintenance Agreement (PMA).
5.1.5
Replacement of Fuses
Hot Surface
The cooling flange can reach a very high temperature.
Warning
Before installing or replacing the fuses, disconnect incoming
mains supply.
When replacing the fuses, follow the steps below:
1.
Disconnect power supply to the instrument.
EF027a