32
CLEANING DISK
1. Open the stopper on the
HypersonIQ cartridge (medication
reservoir).
2. Fill the reservoir half-way with ~3ml
distilled water. Close the stopper.
3. Gently raise the Mouthpiece with
your index finger.
4. Press and hold the power button for
5 seconds. Blue AND yellow indicator
lights appear. Flyp will produce a
pulsing aersosol.
5. Allow the aerosol mist to flow from
the raised Mouthpiece for 1-2 minutes
or until empty.
6. Power Flyp off by pressing the
power button.
7. Remove HypsersonIQ cartridge and
shake out excess water.
8. Let the main unit, cartridge, and
mouthpiece dry on a clean, dry
towel.
9. Allow the interior to dry fully prior to
using again.
It is important to clean the
HypersonIQ disk after each use
in order to prevent the buildup of
residual medication. If these steps are
not followed, the disk may become
blocked – impacting performance.
Please see troubleshooting on page
36 in case of a suspected blockage.