TROUBLE
POSSIBLE CAUSE
CORRECTIVE ACTION
The
nebulization
rate is
extremely low.
The battery is low.
Charge Flyp using a USB port.
Refer to pages 22-23.
An air bubble
has formed in the
medication’s pathway.
Tap the back of the Main
Unit with your index finger
until normal flow resumes.
The Nebulizer Disk has
become clogged with
medication.
See deep cleaning
instructions.
Refer to page 33.
The On-Off
Indicator
Light is on,
but there is no
mist.
The Medication
Reservoir is not correctly
inserted.
Remove the Medication
Reservoir and install it
correctly.
The Medication
Reservoir is not filled.
Fill with medication. Refer to
pages 24-25.
Check that Flyp has
been properly cleaned.
See deep cleaning instructions.
Refer to page 33.