52
Recommended Separation Distances for the Flyp fn2000m
The Flyp™ fn2000m is intended for use in the electromagnetic environment
in which radiated disturbances are controlled. The customer or user of the
Flyp fn2000m can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF Communications
Equipment and the Flyp fn2000m as recommended below, according to the
maximum output power of the communications equipment.
Max Output
Power
(Watts)
Separation (m)
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
Separation (m)
80 to 800MHz
D=(3.5/E1)(Sqrt P)
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(Sqrt P)
0.01
0.116667
0.116667
0.233333
0.1
0.368932
0.368932
0.737865
1
1.166667
1.166667
2.333333
10
3.689324
3.689324
7.378648
100
11.66667
11.66667
23.33333