Guidance and Manufacturer’s Declaration – Immunity
The Flyp fn2000m is intended for use in the electromagnetic environment
specified below. The customer or user of the Flyp fn2000m should ensure that it
is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment – Guidance
ESD:
IEC61000-4-2
±8kV Contact
±2KV, ±4kV,
±8kV, ±15kV Air
±8kV Contact
±2KV, ±4kV,
±8kV, ±15kV Air
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be 30%.
EFT:
IEC61000-4-4
+/- 0.5 KV L1
and L2
+/- 1 KV L1
and L2
+/- 0.5 KV L1
and L2
+/- 1 KV L1
and L2
Mains power quality should be
that of a typical commercial or
hospital environment.
Surges:
IEC61000-4-5
+/- 0.5 KV L1
and L2
+/- 1 KV L1 and L2
+/- 0.5 KV L1
and L2
+/- 1 KV L1 and L2
Mains power quality should be
that of a typical commercial or
hospital environment.