Going off-hook on a line carrying a high dc voltage can 
damage the power supply feeding the line. The test set 
will prevent this from occurring. Unlike the data 
lockout, there is no provision for the operator to 
override a high voltage lockout. Lines with voltages 
exceeding 140 Vdc do not carry analog voice services. If 
you encounter a high voltage lockout, switch the Talk/
Monitor switch back to Monitor (

M

) and carefully 

remove the test leads from the line.

Low Voltage Lockout

Each time the test set is switched to Talk mode, it tests 
the line for high-speed data signals and it measures the 
DC voltage on the line before it goes off-hook. If there 
is less than 3 V on the line, there is not enough voltage 
for the test set to go off-hook and the LCD will display 

LOW VOLT LOCKOUT, DOWN ARROW FOR TALK 
BATTERY

The Low Voltage Lockout message indicates that there 
is very little or no voltage on the telephone line — not 
enough voltage for the test set to perform off-hook 
operations. The most common cause of this message is 
when the test set is switched to Talk mode and it is 
either not connected to anything or it is connected to 
an un-powered pair of wires (dry loop). In this state you 
have the option of activating the test set’s Talk Battery 
function.

There is one additional line condition that should be 
mentioned. When the test set, connected to a line 
carrying data signals but no DC voltage, is switched to 
Talk mode, the LCD will display 

DATA DETECTED AND 

LOW VOLT LOCKOUT

. Some data services have no 

associated DC voltage. The low impedance of the 
TS25D Test Set’s off-hook state would corrupt these 
types of services. Thus the test set does not have any 
provision to override the lockout when this line 
condition is encountered.

In any of the test set’s lockout states, removing the test 
set from the line condition that is causing the lockout 
automatically clears the lockout condition. 

Talk Battery

The TS25D Test Set can power a dry telephone wire pair 
to enable voice communications between itself and 
another test telephone connected to the other end of 
the pair. After turning on Talk Battery at your end, the 
person at the other end should take his test set off-
hook. If the test set at the other end has talk battery 
capability, it should not be enabled. 

The Talk Battery function provides the following: 

Communications capability on 26 gauge wire pairs 
up to about 1000 feet, longer on larger gauges. 

A continuity test. 

When Talk Battery is active, the LCD displays 

LINE = 

SHORTED

 if the test set is connected to a short circuit of 

Ω

 to about 100 

Ω

To enable Talk Battery: 

1

Connect the test set to a dry pair.

2

Set the slide switch to the Talk (

T

) position.

3

Press the down arrow when the message 

LOW 

VOLT LOCKOUT, DOWN ARROW FOR TALK 
BATTERY

 appears.

Note

The Talk Battery function drains the 9 V battery 
more quickly than other TS25D Test Set 
functions.

Содержание TS 25D

Страница 1: ...TS 25D Test Set PN 2451820 October 2005 Rev 3 3 13 2005 2009 2012 2013 Fluke Corporation All rights reserved Printed in China All product names are trademarks of their respective companies Users Guide...

Страница 2: ...or replacement of a defective product which is returned to a Fluke Networks authorized service center within the warranty period To obtain warranty service contact your nearest Fluke Networks authori...

Страница 3: ...des and Functions 6 Monitor Mode 6 Talk Mode 6 Data Lockout 6 Data Lockout Override 7 High Voltage Lockout 7 Low Voltage Lockout 8 Talk Battery 8 Caller ID Call Waiting ID 9 Toner 9 DTMF Detection 10...

Страница 4: ...Pause Duration 14 Flash Duration 14 System Timeout 15 Speaker Timeout 15 Data Lockout 15 Firmware Version 15 Factory Settings 16 Maintenance 16 Keypad Care 16 Replacing the Battery 16 Replacing the L...

Страница 5: ...ccess to valuable information on product updates troubleshooting tips and other support services To register fill out the online registration form on the Fluke Networks website at www flukenetworks co...

Страница 6: ...Flash Memory Dialing Tone and Pulse Dialing Displays Voltage when on hook Displays Current when off hook Polarity Indication Detect and Display Caller ID Call Waiting ID CIDCW Toner Talk Battery DTMF...

Страница 7: ...ric keypad is used to dial telephone numbers and make function selections The keypad includes 12 standard dialing keys the asterisk key and pound key SPKR The SPKR speaker key turns the speaker on inc...

Страница 8: ...of the test set The line cord included with the standard TS25D Test Set has angled bed of nails clips The line cord included with the TS25D CO Test Set has a 346A plug Extension line cords for the 346...

Страница 9: ...the line is not powered Battery The battery icon indicates the battery s charge level The battery level is continuously displayed in Monitor and Talk mode On off hook The telephone icon tells you if...

Страница 10: ...e test set will go off hook and draw dial tone An off hook TS25D test Set works like a standard telephone When off hook the test set is used to perform its main purpose which is to verify the proper o...

Страница 11: ...ARROW TO OVERRIDE Now if you want to go off hook to access the analog phone service sharing the line with ADSL press the down arrow as instructed on the LCD The test set will override the lockout wil...

Страница 12: ...oned When the test set connected to a line carrying data signals but no DC voltage is switched to Talk mode the LCD will display DATA DETECTED AND LOW VOLT LOCKOUT Some data services have no associate...

Страница 13: ...or arrange for someone to call the line the test set is connected to When the call comes in you will hear the Call Waiting Alert tone This will be followed by the CAS tone and the Caller ID message W...

Страница 14: ...ck to Monitor mode DTMF detection is only available when the TS25D Test Set is in Monitor mode Call and Dialing Functions Note Remove the protective film that covers the LCD display before using your...

Страница 15: ...erisk and pound keys are pressed in Pulse mode they will be ignored Memory Dialing The TS25D Test Set allows the user to dial pre programmed numbers in either Tone or Pulse mode If dialing in Pulse mo...

Страница 16: ...slash quote period or 1 depending on how many times you press the key Use the Up arrow to erase characters In this case the Up arrow moves the cursor to the left 7 After the entry is complete press th...

Страница 17: ...rsor location A comma is used to represent a pause Pauses are used to put time delays into a string of numbers The default pause duration is 4 seconds 10 After the entry is complete press the SEL key...

Страница 18: ...screen see Figure 2 Timing Use the Timing option to set the Pause duration and the Flash duration Pause Duration To set the Pause duration 1 Set the slide switch to the Talk T or Monitor M position 2...

Страница 19: ...ss the SEL key or press 3 on the keypad 3 Press 2 SPKR TIMEOUT 4 Highlight the desired time with the arrow keys then press the SEL key or press one of the following keys on the keypad 1 2 minutes TS25...

Страница 20: ...imilar chlorinated solvent on the test set Doing so will damage the test set If moisture should get inside the test set let the test set dry at normal room temperature for 24 hours DO NOT HEAT THE TES...

Страница 21: ...r 5 Loosen the two screws that hold the line cords to the PCB 6 Using needle nose pliers or one of the line cord clips remove the two screws and washers from the line cord connectors 7 Position the sc...

Страница 22: ...last number redial memory Digit Capacity 23 digits per memory Pause Duration User programmable Default 4 seconds Hook Flash Duration User programmable Default 600 ms Automatic Speaker Shut Off Durati...

Отзывы: