SPOT Light
Users Manual
14
General Specifications
Temperature
Operating ............................................................ 10
°
C to 40
°
C (50
°
F to 104
°
F)
Storage ............................................................... -20
°
C to +60
°
C (-4
°
F to +140
°
F)
Humidity
................................................................. 10 % to 90 % non-condensing
Altitude
................................................................... 3,000 m (9,843 ft)
Size (W x H x D)
..................................................... 12.53 cm x 14.86 cm x 4.77 cm (4.94 in x 5.85 in x 1.88 in)
Display
................................................................... LCD Monochrome display
Communication
(USB Device Virtual COM Port) .. Mini B connector for service upload of firmware
Power
..................................................................... Lithium-Ion rechargeable, 3.7 V, 10.75 Wh battery, 2900 mAh
Battery Charger .....................................................
100 V to 240 V, 50/60 Hz input, 6V/3A output. For best performance, battery charger
must be connected to a properly grounded ac receptacle.
Battery Life
............................................................ 10 hours (minimum)
Weight
.................................................................... 0.29 kg (0.7 lb)
Safety .....................................................................
IEC 61010-1: 300 V CAT II, Pollution Degree 2
Ingress Protection Rating
.................................... IEC 60529: IP20
Electromagnetic Environment
............................. IEC 61326-1: Basic
Detailed Specifications
Oximeter SpO
2
Optical Emitter and Detector
%O
2
O
2
Saturations ..................................................... 70 %, 75 %, 80 %, 85 %, 90 %, 95 %, 97 %, 98 %, 99 %, and 100 %
Accuracy
With oximeter manufacturer’s R-curve
Saturation within UUT specific range ..........
±
(1 count + specified accuracy of the UUT)
Saturation outside UUT specific range ....... monotonic with unspecified accuracy
Содержание SPOT Light
Страница 2: ......
Страница 10: ...SPOT Light Users Manual iv...
Страница 12: ...SPOT Light Users Manual vi...
Страница 19: ...SpO2 Functional Tester How to Use the Product 7 gvh006 png Figure 4 Oximeter Sensor Placement...
Страница 24: ...SPOT Light Users Manual 12 gvh002 png Figure 6 External Battery Charger Connections...
Страница 28: ...SPOT Light Users Manual 16...