
SpO2 Functional Tester
Unpack the Product
3
•
Replace the batteries when the low
battery indicator shows to prevent
incorrect measurements.
•
Carefully read all instructions.
•
Do not use the Product around explosive
gas, vapor, or in damp or wet
environments.
•
Do not use, and disable the Product if it
is damaged.
•
Do not use the Product if it operates
incorrectly.
•
Use only current probes, test leads, and
adapters supplied with the Product.
Caution
The pulse oximeter component of the device
is not intended to validate the SpO2 accuracy
of pulse oximeter equipment.
This device is not intended to confirm the
SpO2 accuracy of the calibration curve of the
pulse oximeter monitor or to assess the
optical characteristics of representative
pulse oximeter sensors to determine their
proper calibration.
Not all functional testers and pulse oximeter
equipment are compatible. Functional testers
can vary in pulse methods, pulse contours,
and amplitude. A functional tester might not
accurately reproduce the calibration of the
pulse oximeter equipment and can yield
different results between pulse oximeter
equipment.
Unpack the Product
Carefully unpack all items from the box and check that
you have these items:
•
SPOT Light
•
Users Manual
•
Carrying Case
•
Power Cord
•
AC/DC Power Supply
Содержание SPOT Light
Страница 2: ......
Страница 10: ...SPOT Light Users Manual iv...
Страница 12: ...SPOT Light Users Manual vi...
Страница 19: ...SpO2 Functional Tester How to Use the Product 7 gvh006 png Figure 4 Oximeter Sensor Placement...
Страница 24: ...SPOT Light Users Manual 12 gvh002 png Figure 6 External Battery Charger Connections...
Страница 28: ...SPOT Light Users Manual 16...