SPOT Light
Users Manual
10
Maintenance
The Product is an SpO2 functional tester. Try to prevent
mechanical abuse that could change the test values. The
Product has no internal user-serviceable parts.
Warning
For safe operation and maintenance of the
Product and to prevent personal injury:
•
Repair the Product before use if the
battery leaks.
•
Remove batteries to prevent battery
leakage and damage to the Product if it
is not used for an extended period.
•
Connect the battery charger to the mains
power outlet before the Product.
•
Use only Fluke approved power adapters
to charge the battery.
•
Do not short the battery terminals
together.
•
Do not keep cells or batteries in a
container where the terminals can be
shorted.
•
Keep cells and battery packs clean and
dry. Clean dirty connectors with a dry,
clean cloth.
•
Batteries contain hazardous chemicals
that can cause burns or explode. If
exposure to chemicals occurs, clean
with water and get medical aid.
•
Do not put battery cells and battery
packs near heat or fire. Do not put in
sunlight.
•
Do not disassemble or crush battery
cells and battery packs.
•
Have an approved technician repair the
Product.
•
Use only specified replacement parts.
•
Remove the input signals before you
clean the Product.
•
Connect factory supplied three-
conductor mains power cord to a
grounded power outlet.
•
Do not use a two-conductor adapter or
extension cord.
Содержание SPOT Light
Страница 2: ......
Страница 10: ...SPOT Light Users Manual iv...
Страница 12: ...SPOT Light Users Manual vi...
Страница 19: ...SpO2 Functional Tester How to Use the Product 7 gvh006 png Figure 4 Oximeter Sensor Placement...
Страница 24: ...SPOT Light Users Manual 12 gvh002 png Figure 6 External Battery Charger Connections...
Страница 28: ...SPOT Light Users Manual 16...