SPOT Light
Users Manual
2
Table 1. Symbols
Symbol
Description
Important information. Refer to manual.
Hazardous voltage.
Conforms to relevant Canadian and US
standards.
Conforms to relevant Australian EMC
requirements.
Conforms to European Union directives.
Do not dispose of this product as
unsorted municipal waste. Go to Fluke’s
website for recycling information.
Intended Use
The Product is intended to be used to test and verify the
basic operation of patient monitoring devices or systems
used to monitor SpO2. Additionally, the Product provides
an optical signal to verify the electronics inside the pulse
oximeter sensor are functional.
The intended user is a trained biomedical equipment
technician who performs periodic preventative
maintenance checks on patient monitors in service. Users
can be associated with hospitals, clinics, original
equipment manufacturers and independent service
companies that repair and service medical equipment.
The end user is an individual, trained in medical
instrumentation technology.
This Product is not intended for use on patients, or to test
devices while connected to patients. This Product is not
intended to be used to calibrate medical equipment.
Safety Information
A
Warning
identifies conditions and procedures that are
dangerous to the user. A
Caution
identifies conditions
and procedures that can cause damage to the Product or
the equipment under test.
Warning
To prevent possible electrical shock, fire, or
personal injury, follow these guidelines:
•
Do not connect the Product to a patient
or equipment connected to a patient. The
Product is intended for equipment
evaluation only and should never be
used in diagnostics, treatment, or any
other capacity where the Product would
come in contact with a patient.
•
Use the Product only as specified, or the
protection supplied by the Product can
be compromised.
Содержание SPOT Light
Страница 2: ......
Страница 10: ...SPOT Light Users Manual iv...
Страница 12: ...SPOT Light Users Manual vi...
Страница 19: ...SpO2 Functional Tester How to Use the Product 7 gvh006 png Figure 4 Oximeter Sensor Placement...
Страница 24: ...SPOT Light Users Manual 12 gvh002 png Figure 6 External Battery Charger Connections...
Страница 28: ...SPOT Light Users Manual 16...