Put the sensor with the LEDs on the bottom of the 

artificial finger. While you put the sensor on the artifact 

finger, monitor the signal indicator along the bottom of the 

Product display. Adjust the sensor on the finger for 

maximum signal strength. 

Note 

Cable position changes with sensor 

manufacturer 

 

How to Set Test Parameters 

When you turn on the Product, all parameters are set to 

their default values. To change a parameter value, push 

 or 

 to move the highlight to a parameter that 

you need to change. Push 

 to move the highlight to 

the value of the parameter. Push 

 or 

 to 

scroll through the values. With the correct parameter 

value shown in the display, push 

 to set the 

parameter. Table 5 is a list of all parameters and their 

values. 

Table 5. Test Parameters 

Parameter 

Values* 

SpO2 

70%, 75%, 80%, 85%, 90%, 

95%, 

97%

, 98%, 99%, 100% 

HR (Heart Rate) 

30, 40, 60, 65, 

80

, 100, 120, 

140, 150, 180, and 240 BPM 

PA (Pulse Amplitude)  0.2%, 

2.0%

, and 10% 

Transmission 

LG (Large), 

Med (Medium),

 

and Sm (Small) finger 

Artifact 

None

, Respiration: 2.5%, 

Ambient light: 50 or 60 Hz 

Type 

Nonin, Masimo, 

Nellcor

, Nihon 

Kohden, Mindray, GE-Ohmeda, 

Philips/HP, and BCI 

Test 

Manual

, Custom 1, Custom 2, 

and Custom 3 

* Default values are in bold type. 

 

Содержание SPOT Light

Страница 1: ...Rev 2 1 22 2012 2022 Fluke Corporation All rights reserved Printed in USA Specifications are subject to change without notice All product names are trademarks of their respective companies SPOT Light...

Страница 2: ......

Страница 3: ...TNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA ARISING FROM ANY CAUSE O...

Страница 4: ...Notify the carrier and ask for an agent to be present while the instrument is unpacked There are no special unpacking instructions but be careful not to damage the instrument when unpacking it Inspec...

Страница 5: ...le we recommend the following guide for repackaging Use a double walled carton of sufficient strength for the weight being shipped Use heavy paper or cardboard to protect all instrument surfaces Use n...

Страница 6: ...sult in electrical shock hazards or improper operation Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications Restrictions and Liabilities Info...

Страница 7: ...k the Product 3 Instrument Familiarization 4 Accessories 5 How to Turn On the Product 6 How to Use the Product 6 SpO2 Sensor Placement 6 How to Set Test Parameters 8 How to Set Custom Tests 9 Maintena...

Страница 8: ...SPOT Light Users Manual ii Battery Removal 13 General Specifications 14 Detailed Specifications 14 Oximeter SpO2 Optical Emitter and Detector 14...

Страница 9: ...iii List of Tables Table Title Page 1 Symbols 2 2 Product Controls and Connections 4 3 Standard Accessories 5 4 Optional Accessories 5 5 Test Parameters 8...

Страница 10: ...SPOT Light Users Manual iv...

Страница 11: ...res Figure Title Page 1 Product Controls and Connections 4 2 Power Up Screen 6 3 Main Screen 6 4 Oximeter Sensor Placement 7 5 SpO2 Parameter Change 9 6 External Battery Charging Connections 12 7 Batt...

Страница 12: ...SPOT Light Users Manual vi...

Страница 13: ...luke Biomedical SPOT Light SpO2 Functional Tester the Product is a compact portable functional tester used to measure the performance of SpO2 monitors pulse oximeters The Product uses light detection...

Страница 14: ...ls clinics original equipment manufacturers and independent service companies that repair and service medical equipment The end user is an individual trained in medical instrumentation technology This...

Страница 15: ...ulse oximeter equipment This device is not intended to confirm the SpO2 accuracy of the calibration curve of the pulse oximeter monitor or to assess the optical characteristics of representative pulse...

Страница 16: ...controls and connections shown in Figure 1 gvh003 png Figure 1 Product Controls and Connections Table 2 Product Controls and Connections Item Description 1 SpO2 Artificial Finger 2 Scroll Up and Down...

Страница 17: ...Biomedical Part Number SPOT Light Users Manual 4151274 AC DC Power Supply 4766589 AC Power Cord US 284174 Schuko 769422 UK 769455 Japan 284174 Australia China 658641 Brazil 1 3841347 Carrying Case 402...

Страница 18: ...shown is for illustration only The version shown on your Product could be different When the self test is complete and no errors are sensed the screen shown in Figure 3 shows in the display gvh004 jpg...

Страница 19: ...SpO2 Functional Tester How to Use the Product 7 gvh006 png Figure 4 Oximeter Sensor Placement...

Страница 20: ...arameter that you need to change Push to move the highlight to the value of the parameter Push or to scroll through the values With the correct parameter value shown in the display push to set the par...

Страница 21: ...ustom Tests When the Test parameter is set to Manual no parameter values are stored A maximum of three custom tests can be stored in the Product To set up a custom test 1 Push or to move the highlight...

Страница 22: ...tery Do not short the battery terminals together Do not keep cells or batteries in a container where the terminals can be shorted Keep cells and battery packs clean and dry Clean dirty connectors with...

Страница 23: ...battery message shows in the display How to Charge the Battery The battery charge level is shown in the lower right corner of the display when the battery pack is installed in the Product If the batt...

Страница 24: ...SPOT Light Users Manual 12 gvh002 png Figure 6 External Battery Charger Connections...

Страница 25: ...emove the battery pack 1 Push down on the battery pack latch as shown in Figure 7 2 Pull the battery pack from the Product gvh023 png Figure 7 Battery Removal To put the battery pack into the Product...

Страница 26: ...tery Charger 100 V to 240 V 50 60 Hz input 6V 3A output For best performance battery charger must be connected to a properly grounded ac receptacle Battery Life 10 hours minimum Weight 0 29 kg 0 7 lb...

Страница 27: ...sed in parts per million ppm Ratios Large finger 12 00 ppm medium finger 80 00 ppm and small finger 300 00 ppm Accuracy 50 30 for compatible monitors unspecified for others Selected by finger size and...

Страница 28: ...SPOT Light Users Manual 16...

Отзывы: