RUSKA 7252
Users Manual
4-28
Menu | Test
The
Menu | Test
section of the system is used to perform a number of the diagnostic
and tuning functions that are available with the RUSKA 7252. The Sweep and Control
test operate on the current selected channel. The other screens are independent of the
selected channel.
glb37.bmp
Figure 4-25. Channel A: Menu | Test - Menu
Menu | Test — Sweep Test
The sweep function can be used to automatically exercise the elastic sensing element of
the device under test prior to performing a calibration. For the Calibrator to perform the
sweep test, input the high and low set-point pressures, the pressure control tolerance, the
dwell time at set-point, and the number of cycles to perform.
1.
Ensure that the units, limits, and control parameters are set to their desired values.
2.
The sweep test is entered from the Sweep Test Menu. From the Main Menu, (press
PREVIOUS
until the Main Menu appears), press
Menu
[
F6
]
| Test
[
F5
]
|
Sweep
[
F2
].
3.
Enter the high and low pressure points, the control tolerance, dwell time, and the
number of cycles.