Summary of Error Messages, cont.
-315
Configuration Data Lost. The configuration data has been lost. Check all parameters to be sure
they are correct.
-330
Self-Test Failed. Check the display for the test that failed.
-350
Queue Overflow. The error queue was full and messages were lost.
500
Controller Malfunction. Internal control failure.
501
High Limit Exceeded. The pressure was greater than the high limit.
502
Low Limit Exceeded. The pressure was less that the low limit.
503
Slew limit Exceeded. The pressure changed faster than the slew limit allowed.
521
Pressure Overrange. The pressure reading is outside the range of the Calibrator.
531
Oven Temp Overrange. Either the transistor that drives the heater for the quartz Bourdon tube
sensor or the oven temperature sensor itself is malfunctioning. To observe the oven temperature,
select OK then select MENU/DISP. Check the transistor and sensor for malfunction, requesting
service if necessary.
533
Case Pressure Overrange. Select OK, then reduce the pressure at the case reference port to
30 psia or lower.
540
Mechanical Zeroing Needed. The zero point of the quartz Bourdon sensor is beyond the range of
the compensation circuit. The zero is adjusted by the software but should be manually adjusted for
complete accuracy.
542
Oven Control Failure. The temperature controller is unable to keep the sensor at the proper
temperature.
545
Sensor Communication Error. Unable to Communicate with 7215
xi
Sensor
546
Sensor Calibration Lost. The 7215xi sensor has lost its calibration and must be recalibrated.
600
Factory Data Lost. Internal factory constants have been lost. Contact Fluke for more information.
601
Calibration Mode. The Calibrate button must be pressed before SCPI calibration commands can
be executed.
800
Solenoid Over-Temperature. The control solenoids have over-heated. Wait until they have cooled
before entering Control mode.
801
Error reading Barometric RPT Pressure Sensor
802
Control Sensor out of range
803
Controller Communication Error
804 Autotune
Failed
Содержание RUSKA 7252
Страница 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Страница 4: ......
Страница 10: ...RUSKA 7252 Users Manual vi ...
Страница 12: ...RUSKA 7252 Users Manual viii ...
Страница 20: ...RUSKA 7252 Users Manual 1 6 ...
Страница 40: ...RUSKA 7252 Users Manual 3 6 ...
Страница 72: ...RUSKA 7252 Users Manual 4 32 ...
Страница 114: ...RUSKA 7252 Users Manual 6 22 ...
Страница 118: ...RUSKA 7252 Users Manual 7 4 ...