Summary of Error Messages
Summary of Error Messages
Negative error numbers are from the Standard Commands for Programmable Instruments
(Version 1991.0).
Value
Description and Corrective Action
0 No
Error.
-103
Invalid Separator. Check punctuation in the SCPI command.
-104
Data Type. The type of parameter data is incorrect.
-109
Missing Parameter. No valid parameter was found for the SCPI command.
-110
Command Header. The command name is not valid.
-113
Command Unknown. The command specified does not exist.
-114
Header Suffix. The numeric suffix for the command name is out of range.
-221
Settings Conflict. The command could not be executed due to the current state of the Calibrator.
Some commands cannot be executed while a program, self-test, or calibration is in progress.
-222
Out of Range. The value is not within the valid range. For pressures check high and low limits.
-281
Cannot create program. Program memory is full.
-282
Illegal Program Name. The name specified is not valid or does not exist.
-284
Program Currently Running. The command cannot be executed while a program is running.
-285
Program Syntax Error. The syntax of the program definition is not correct.
-286
Program Runtime Error. An error occurred while running the program. Usually the setpoint is out of
range.
-313
Calibration Data Lost. The calibration data has been lost and the unit must be recalibrated.
Содержание RUSKA 7252
Страница 3: ...7252 Change Language Hold key for 5 seconds enter mode ...
Страница 4: ......
Страница 10: ...RUSKA 7252 Users Manual vi ...
Страница 12: ...RUSKA 7252 Users Manual viii ...
Страница 20: ...RUSKA 7252 Users Manual 1 6 ...
Страница 40: ...RUSKA 7252 Users Manual 3 6 ...
Страница 72: ...RUSKA 7252 Users Manual 4 32 ...
Страница 114: ...RUSKA 7252 Users Manual 6 22 ...
Страница 118: ...RUSKA 7252 Users Manual 7 4 ...