![Fluke ProcessMeter 787 Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/fluke/processmeter-787/processmeter-787_calibration-manual_2309971032.webp)
Model 787
Calibration Manual
24
OFF
A
mA
COM
V
1000V
30mA
FUSED
0.44A
(1A /30 sec)
FUSED
mA
mA
A
mA
OUTPUT 0-24mA
SOURCE SIMULATE
+
+
% STEP
COARSE
FINE
mV
V
V
OUTPUT
CAT
787
PROCESSMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
UUT
5500A
POWER
I
O
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
EARTH
SCOPE
BOOST
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
A
MULT
x
DIV
÷
OUT
TRIG
5500A CALIBRATOR
20V PK
MAX
HI
LO
TC
TRIG
OUT
1000V
RMS
MAX
20V
RMS
MAX
1V PK
MAX
20V PK
MAX
NORMAL
AUX
SCOPE
V, ,
RTD
A, -SENSE,
AUX V
200V PK
MAX
STBY
(BLUE)
LT0009F.EPS
Figure 10. AC/DC Voltage Measurement Test Connections
Table 10. AC Volts Test
787 Range
Calibrator Voltage
and Frequency
Minimum Acceptable
Reading
Maximum Acceptable
Reading
400 mV ac
100 mV @ 60 Hz
98.9 mV
101.1 mV
400 mV ac
300 mV @ 60 Hz
297.5 mV
302.5 mV
4 V ac
1 V @ 60 Hz
0.991 V
1.009 V
4 V ac
2 V @ 60 Hz
1.984 V
2.016 V
4 V ac
3 V @ 60 Hz
2.977 V
3.023 V
40 V ac
10 V @ 60 Hz
9.91 V
10.09 V
40 V ac
30 V @ 60 Hz
29.77 V
30.23 V
400 V ac
100 V @ 60 Hz
99.1 V
100.9 V
400 V ac
300 V @ 60 Hz
297.7 V
302.3 V
1000 V ac
100 V @ 60 Hz
97
103
1000 V ac
800 V @ 60 Hz
792
808