Calibration Manual
Specifications
7
Ohms
Range
Resolution
Measurement Current
Accuracy
±
(% of R Counts)
400.0
Ω
0. 1
Ω
220
µ
A
0.2 % + 2
4.000 k
Ω
0.001 k
Ω
59
µ
A
0.2 % + 1
40.00 k
Ω
0.01 k
Ω
5.9
µ
A
0.2 % + 1
400.0 k
Ω
0.1 k
Ω
590 nA
0.2 % + 1
4.000 M
Ω
0.001 M
Ω
220 nA
0.35 % + 3
40.00 M
Ω
0.01 M
Ω
22 nA
2.5 % + 3
Overload protection: 1000 V
Open circuit voltage: <3.9 V
Frequency Counter Accuracy
Range
Resolution
Accuracy
±
(% of R Counts)
199.99 Hz
0.01 Hz
0.005 % + 1
1999.9 Hz
0.1 Hz
0.005 % + 1
19.999 kHz
0.001 kHz
0.005 % + 1
Display updates 3 times/second at >10 Hz
Frequency Counter Sensitivity
Input Range
Minimum Sensitivity (rms Sinewave)
0.5 Hz to 20 kHz*
1 V
0.1 V
4 V
1 V
40 V
3 V
400 V
30 V
1000 V
300 V
* Usable 0.5 Hz to 20 kHz with reduced sensitivity
Diode Test and Continuity Test
Diode test indication
Display voltage drop: 0.2 mA nominal test current at 0.6 V.
2.4 V full scale, accuracy
±
( 2 % + 1 count)
Continuity test indication
continuous audible tone for test resistance <100
Ω
Open circuit voltage
<3.9 V
Short circuit current
1.2 mA typical
Overload protection
1000 V rms