Calibration Manual
Specifications
5
Table 2. Required Equipment and Software
Equipment
Minimum Specifications
Recommended Model
Calibration Source
No Substitute
Fluke Model 5500A, 5520A
Digital Multimeter
No Substitute
HP 3458A, HP 34401A*
Test Leads, low leakage,
RG-58/U type
Leakage resistance
>
than 1.0 x 10
13
Ω
at 45
°
C and 75 % relative humidity
Fluke 5440A-7002 Low Thermal
Test Leads
Calibration Module
w/Battery Eliminator
W
No Substitute. A generic battery
eliminator might pose a safety
hazard. Includes Cal Module,
Battery Eliminator, Calibration
Manual, and MET/CAL Calibration
Software (provided on floppy disk).
Fluke-787CAL 120 (USA/Canada)
Fluke-787CAL 230 (Europe)
Fluke-787CAL 240 (UK)
Fluke-787CAL 240AN (Australia)
Fluke-787CAL 100 (Japan)
Battery Eliminator,
Input: 120-240 V AC;
Output: 12 V DC/ 300 mA.
No Substitute. Only use a BE860
battery eliminator.
Fluke BE860 120 (in USA/Can)
Fluke BE860 230 (Europe)
Fluke BE860 240 (UK)
Fluke BE860 240AN (Aus)
Null Modem Cable
9-pin to 9-pin
Fluke RS 43
Personal Computer
Windows
3.1 or later (Refer to
MET/CAL
manual set), 16 MB
RAM, 80 MB on disk, VGA monitor.
486 (or later) IBM-compatible
Calibration Software
No Substitute
Fluke
MET/CAL Calibration
Software 6.0 or later
* HP 34401A may only be used in conjuction with Fluke 5520A.
Specifications
The 787 ProcessMeter’s performance and accuracy are specified for one year after
calibration, at operating temperatures of +18
°
C to +28
°
C (64
°
F to 82
°
F), in relative
humidity to 90 %, after the ProcessMeter has been warmed up for 5 minutes.
Note
A “count” is the amount by which the least significant digit can vary.
DC Volts
Range (V dc)
Resolution
Accuracy,
±
(% of R Counts)
4.000
0.001 V
0.1 % + 1
40.00
0.01 V
0.1 % + 1
400.0
0.1 V
0.1 % + 1
1000
1 V
0.1 % + 1
Input impedance: 10 M
Ω
(nominal), < 100 pF
Normal mode rejection ratio: >60 dB at 50 Hz or 60 Hz
Common mode rejection ratio: >120 dB at dc, 50 Hz, or 60 Hz
Overvoltage protection: 1000 V