Using the Loss/Length Option
Loss/Length Test Settings
7-7
7
Table 7-1. Loss/Length Test Settings (cont.)
Loss/Length tab
TEST METHOD
Loss results include connections added after referencing. The reference and test
connections determine which connections are included in results. The Test Method
refers to the number of end connections included:
2 Jumper: Loss results include one connection at one end of the link.
1 Jumper: Loss results include connections at both ends of the link. The reference
and test connections shown in this Chapter produce 1 jumper results. See "About
1 Jumper Connections" on page 7-10.
3 Jumper: Loss results exclude connections at the ends of the link. Only the fiber
loss is measured.
Different standards have different names for the three test methods. See
Appendix A for details.
The TEST METHOD setting does not affect loss results. It is only saved with the
results to record which method you used. This setting does affect the reference
and test diagrams shown on the tester's display. The diagrams show connections
for the method selected.
CONNECTOR
TYPE
Select the type of connector used in the cabling. This setting affects the diagrams
shown for reference connections. If the cabling's connector type is not listed, use
General.
-continued-
SETUP Tab
Setting
Description
Содержание OF-500-01 OptiFiber
Страница 4: ......
Страница 18: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook xiv...
Страница 29: ...Getting Acquainted Powering the Tester 1 9 1 ajt20f eps Figure 1 1 Battery Pack Feature...
Страница 31: ...Getting Acquainted Verifying Operation 1 11 1 ajt56f eps Figure 1 2 Removing the Module...
Страница 48: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 1 28...
Страница 76: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 2 28...
Страница 95: ...Using the OTDR Running the OTDR Test 3 19 3 ajt33f eps Figure 3 12 Connecting the OTDR to Spooled Cable...
Страница 110: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 34...
Страница 117: ...Using the OTDR Comparing OTDR Traces 3 41 3 ajt68f eps Figure 3 16 Comparing OTDR Traces F G H C D E A B...
Страница 132: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 3 56...
Страница 135: ...Using the ChannelMap Function Running the Test 4 3 4 ajt55f eps Figure 4 2 ChannelMap Test Connections...
Страница 138: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 4 6...
Страница 141: ...Using the FaultMap Function Running the Test 5 3 5 ajt55f eps Figure 5 2 FaultMap Test Connections Channel to be tested...
Страница 144: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 5 6...
Страница 155: ...Using the Loss Length Option Changing the Connector Adapter 7 3 7 ajt61f eps Figure 7 2 Changing the Connector Adapter...
Страница 196: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 7 44...
Страница 200: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 8 4 ajt03f eps Figure 8 2 Using the Visual Fault Locator...
Страница 216: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 10 10...
Страница 254: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 12 30...
Страница 256: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook A 2...
Страница 259: ...Loss Test Methods 1 Jumper Method B 3 B ajt59f eps Figure B 1 1 Jumper Reference and Test Connections singlemode shown...
Страница 261: ...Loss Test Methods 2 Jumper Method B 5 B ajt58f eps Figure B 2 2 Jumper Reference and Test Connections singlemode shown...
Страница 268: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook C 2...
Страница 278: ...OF 500 OptiFiber Certifying OTDR Technical Reference Handbook 10 OTDR 3 11 WEAK SIGNAL 3 35 Z TEST METHOD B 8 zoom 3 27...