9640A
Instruction Manual
5-28
Note
Ultra-Low level measurements are made relative to levels previously
measured with the measuring receiver. Two levels (-85dBm and -95dBm)
are used as reference points to provide the required UUT hardware
configurations for the subsequent tests. The value of the output level at -
85dBm and -95dBm measured in steps 30 through 37 above will be used as
references in the following procedure, identified as P
-85
and P
-95
respectively for each test frequency. It is important that the correct
sequence is followed as described.
Making these precision ultra-low level measurements requires appropriate
choice of measuring receiver settings to provide adequate linearity, noise
floor, and reading repeatability throughout the amplitude range required.
For any given frequency the measuring receiver the settings must be
maintained for all amplitudes at that frequency after establishing the
reference point. In particular, receiver attenuator, reference level, and
resolution bandwidth settings must not change.
37.
On the UUT enable the REF FREQUENCY OUTPUT at 10 MHz and set the UUT as
follows:
Frequency 10
MHz
Level -85
dBm
Output
OPER
38.
Allow the measuring receiver to tune the input signal, and then set the controls to
measure RF level, relative (dB) to the current input level.
39.
Without changing any measuring receiver settings, set the UUT to the next amplitude
listed in part A of Table 5-15 for this frequency. Allow the measuring receiver
reading to stabilize and record the reading as
P
rel
. (Note that
P
rel
will be a negative
number in dB).
40.
Calculate the UUT output
P
out
=
P
-85
+
P
rel
. Check that the value of
P
out
is within
the tolerance shown in part A of Table 5-15.
41.
Repeat steps 39 and 40 for each amplitude listed in part A of Table 5-15 for this
frequency.
42.
Return the UUT output to -85dBm, set the UUT frequency to the next frequency
listed in part A of Table 5-15 and repeat steps 38 through 41.
43.
Set the UUT to 10 MHz at -95dBm. Repeat steps 38 through 42 for the points listed
in part B of Table 5-15 using the calculation
P
out
=
P
-95
+
P
rel
in step 40.
44.
Set the UUT to
STBY
.
Содержание 9640A Series
Страница 8: ......
Страница 16: ...9640A Instruction Manual viii...
Страница 22: ...9640A Instruction Manual 1 2...
Страница 40: ...9640A Instruction Manual 2 2...
Страница 108: ...9640A Instruction Manual 4A 2...
Страница 138: ...9640A Instruction Manual 4C 2...
Страница 164: ...9640A Instruction Manual 4D 2...
Страница 168: ...9640A Instruction Manual 4D 6...
Страница 170: ...9640A Instruction Manual 4E 2...
Страница 174: ...9640A Instruction Manual 4E 6...
Страница 176: ...9640A Instruction Manual 4F 2...
Страница 184: ...9640A Instruction Manual 4F 10...
Страница 186: ...9640A Instruction Manual 4G 2...
Страница 190: ...9640A Instruction Manual 4G 6...
Страница 192: ...9640A Instruction Manual 5 2...
Страница 272: ...9640A Instruction Manual 5 82...
Страница 274: ...9640A Instruction Manual 6 2...
Страница 284: ...9640A Instruction Manual 7 2...
Страница 309: ...8 1 Chapter 8 Lists of Replaceable Parts Title Page Introduction 8 3 How to Obtain Parts 8 3 Parts Lists 8 4...
Страница 310: ...9640A Instruction Manual 8 2...
Страница 316: ...9640A Instruction Manual 8 8...
Страница 320: ...9640A Instruction Manual A 4...